Cyclic voltammetry

Cyclic voltammetry
Typical cyclic voltammogram where ipc and ipa show the peak cathodic and anodic current respectively for a reversible reaction.

Cyclic voltammetry or CV is a type of potentiodynamic electrochemical measurement. In a cyclic voltammetry experiment the working electrode potential is ramped linearly versus time like linear sweep voltammetry. Cyclic voltammetry takes the experiment a step further than linear sweep voltammetry which ends when it reaches a set potential. When cyclic voltammetry reaches a set potential, the working electrode's potential ramp is inverted. This inversion can happen multiple times during a single experiment. The current at the working electrode is plotted versus the applied voltage to give the cyclic voltammogram trace. Cyclic voltammetry is generally used to study the electrochemical properties of an analyte in solution.[1][2][3]

Contents

Experimental method

cyclic voltammetry waveform

In cyclic voltammetry, the electrode potential ramps linearly versus time as shown. This ramping is known as the experiment's scan rate (V/s). The potential is applied between the reference electrode and the working electrode and the current is measured between the working electrode and the counter electrode. This data is then plotted as current (i) vs. potential (E). As the waveform shows, the forward scan produces a current peak for any analytes that can be reduced (or oxidized depending on the initial scan direction) through the range of the potential scanned. The current will increase as the potential reaches the reduction potential of the analyte, but then falls off as the concentration of the analyte is depleted close to the electrode surface. If the redox couple is reversible then when the applied potential is reversed, it will reach the potential that will reoxidize the product formed in the first reduction reaction, and produce a current of reverse polarity from the forward scan. This oxidation peak will usually have a similar shape to the reduction peak. As a result, information about the redox potential and electrochemical reaction rates of the compounds are obtained.

For instance if the electronic transfer at the surface is fast and the current is limited by the diffusion of species to the electrode surface, then the current peak will be proportional to the square root of the scan rate. This relationship is described by the Cottrell equation.

Characterization

The utility of cyclic voltammetry is highly dependent on the analyte being studied. The analyte has to be redox active within the experimental potential window. It is also highly desirable for the analyte to display a reversible wave. A reversible wave is when an analyte is reduced or oxidized on a forward scan and is then reoxidized or rereduced in a predictable way on the return scan as shown in the first figure.

Even reversible couples contain polarization overpotential and thus display a hysteresis between absolute potential between the reduction (Epc) and oxidation peak (Epa). This overpotential emerges from a combination of analyte diffusion rates and the intrinsic activation barrier of transferring electrons from an electrode to analyte. A theoretical description of polarization overpotential is in part described by the Butler-Volmer equation and Cottrell equation. Conveniently in an ideal system the relationships reduces to, |E_{pc}-E_{pa}|=\frac{57\text{ mV}}{n}, for an n electron process.[2]

Reversible couples will display a ratio of the peak currents passed at reduction (ipc) and oxidation (ipa) that is near unity (1 = ipa/ipc). This ratio can be perturbed for reversible couples in the presence of a following chemical reaction, stripping wave, or nucleation event.

When such reversible peaks are observed thermodynamic information in the form of half cell potential E01/2 can be determined. When waves are semi-reversible such as when ipa/ipc is less than or greater than 1, it can be possible to determine even more information especially kinetic processes like following chemical reaction.

When waves are non-reversible it is impossible to determine what their thermodynamic E01/2 is with cyclic voltammetry. This E01/2 can be determined, however it often requires equal quantities of the analyte in both oxidation states. When a wave is non-reversible cyclic voltammetry can not determine if the wave is at its thermodynamic potential or shifted to a more extreme potential by some form of overpotential. The couple could be irreversible because of a following chemical process, a common example for transition metals is a shift in the geometry of the coordination sphere. If this is the case, then higher scan rates may show a reversible wave. It is also possible that the wave is irreversible due to a physical process most commonly some form of precipitation as discussed below. Some speculation can be made in regards to irreversible waves however they are generally outside the scope of cyclic voltammetry.

Experimental setup

The method uses a reference electrode, working electrode, and counter electrode which in combination are sometimes referred to as a three-electrode setup. Electrolyte is usually added to the test solution to ensure sufficient conductivity. The combination of the solvent, electrolyte and specific working electrode material determines the range of the potential.

Electrodes are static and sit in unstirred solutions during cyclic voltammetry. This "still" solution method results in cyclic voltammetry's characteristic diffusion controlled peaks. This method also allows a portion of the analyte to remain after reduction or oxidation where it may display further redox activity. Stirring the solution between cyclic voltammetry traces is important as to supply the electrode surface with fresh analyte for each new experiment. The solubility of an analyte can change drastically with its overall charge. Since cyclic voltammetry usually alters the charge of the analyte it is common for reduced or oxidized analyte to precipitate out onto the electrode. This layering of analyte can insulate the electrode surface, display its own redox activity in subsequent scans, or at the very least alter the electrode surface. For this and other reasons it is often necessary to clean electrodes between scans.

Common materials for working electrodes include glassy carbon, platinum, and gold. These electrodes are generally encased in a rod of inert insulator with a disk exposed at one end. A regular working electrode has a radius within an order of magnitude of 1 mm. Having a controlled surface area with a defined shape is important for interpreting cyclic voltammetry results.

To run cyclic voltammetry experiments at high scan rates a regular working electrode is insufficient. High scan rates create peaks with large currents and increased resistances which result in distortions. Ultramicroelectrodes can be used to minimize the current and resistance.

The counter electrode, also known as the auxiliary or second electrode, can be any material which conducts easily and won't react with the bulk solution. Reactions occurring at the counter electrode surface are unimportant as long as it continues to conduct current well. To maintain the observed current the counter electrode will often oxidize or reduce the solvent or bulk electrolyte.

Reference electrodes are a complex subject and worth investigating elsewhere.

Variations

In some experiments an electroactive species is fixed to the surface of the electrode, for instance in microparticle voltammetry.

Potentiodynamic techniques also exist that add low-amplitude ac perturbation to a potential ramp and measure variable response in a single frequency (ac voltammetry) or in many frequencies simultaneously (potentiodynamic electrochemical impedance spectroscopy).[4] The response in alternating current is two-dimensional – it is characterised by amplitude and phase. The amplitude and phase depend differently on frequency for constituents of ac response attributed to different processes (charge transfer, diffusion, double layer charging, etc.). Frequency response analysis enables simultaneous monitoring of the various processes that contribute to the potentiodynamic ac response of electrochemical system.

Distinctions

Cyclic voltammetry is not a hydrodynamic technique. In a hydrodynamic technique flow is achieved at the electrode surface by stirring the solution, pumping the solution, or rotating the electrode as is the case with rotating disk electrodes and rotating ring-disk electrodes. These techniques target steady state conditions which appear the same scanned from the positive or the negative, thus limiting them to linear sweep voltammetry.

See also

References

  1. ^ Bard, Allen J.; Larry R. Faulkner (2000-12-18). Electrochemical Methods: Fundamentals and Applications (2 ed.). Wiley. ISBN 0471043729. 
  2. ^ a b Nicholson, R. S.; Irving. Shain (1964-04-01). "Theory of Stationary Electrode Polarography. Single Scan and Cyclic Methods Applied to Reversible, Irreversible, and Kinetic Systems.". Analytical Chemistry 36 (4): 706–723. doi:10.1021/ac60210a007. 
  3. ^ Heinze, Jurgen (1984). "Cyclic Voltammetry-"Electrochemical Spectroscopy". New Analytical Methods (25)". Angewandte Chemie International Edition in English 23 (11): 831–847. doi:10.1002/anie.198408313. 
  4. ^ http://www.abc.chemistry.bsu.by/vi/

Further reading

  • Bard, Allen J.; Larry R. Faulkner (2000-12-18). Electrochemical Methods: Fundamentals and Applications (2 ed.). Wiley. ISBN 0471043729. 
  • Zoski, Cynthia G. (2007-02-07). Handbook of Electrochemistry. Elsevier Science. ISBN 0444519580. 
  • Kissinger, Peter; William R. Heineman (1996-01-23). Laboratory Techniques in Electroanalytical Chemistry, Second Edition, Revised and Expanded (2 ed.). CRC. ISBN 0824794451. 

External links


Wikimedia Foundation. 2010.

Игры ⚽ Поможем сделать НИР

Look at other dictionaries:

  • cyclic voltammetry — ciklinė voltamperometrija statusas T sritis Standartizacija ir metrologija apibrėžtis Voltamperinis matavimas, kai poliarizacinės kreivės nustatomos periodiškai keičiant elektrodo potencialo skleidimo kryptį. atitikmenys: angl. cyclic voltammetry …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • Voltammetry — is a category of electroanalytical methods used in analytical chemistry and various industrial processes. In voltammetry, information about an analyte is obtained by measuring the current as the potential is varied. Three electrode system… …   Wikipedia

  • Differential pulse voltammetry — (AKA Differential Pulse Polarography or DPP) is often used to make electrochemical measurements. It can be considered as a derivative of linear sweep voltammetry or staircase voltammetry, with a series of regular voltage pulses superimposed on… …   Wikipedia

  • Neopolarogram — mathematical relation of current, charge and fractional derivatives. The term Neopolarogram refers to mathematical derivatives of polarograms or cyclic voltammograms that in effect deconvolute diffusion and electrochemical kinetics. This is… …   Wikipedia

  • analysis — /euh nal euh sis/, n., pl. analyses / seez /. 1. the separating of any material or abstract entity into its constituent elements (opposed to synthesis). 2. this process as a method of studying the nature of something or of determining its… …   Universalium

  • Coulometry — is the name given to a group of techniques in analytical chemistry that determine the amount of matter transformed during an electrolysis reaction by measuring the amount of electricity (in coulombs) consumed or produced.[1] There are two basic… …   Wikipedia

  • Potentiometer (measuring instrument) — A potentiometer is an instrument for measuring the potential (voltage) in a circuit. Before the introduction of the moving coil and digital volt meters, potentiometers were used in measuring voltage, hence the meter part of their name. The method …   Wikipedia

  • Phoenix (spacecraft) — This article is about the Mars lander. For the Star Trek spacecraft, see Phoenix (Star Trek). For other uses, see Phoenix (disambiguation). Phoenix Mars Mission Artist s impression of the Phoenix spacecraft as it lands on Mars Operator NASA …   Wikipedia

  • Rotating disk electrode — A rotating disk electrode (RDE) is a hydrodynamic working electrode used in a three electrode system. [Bard, A.J.; Faulkner, L.R. Electrochemical Methods: Fundamentals and Applications. New York: John Wiley Sons, 2nd Edition, 2000.] Structure The …   Wikipedia

  • Hydrodynamic technique — An electroanalytical method in which the analyte solution flows relative to a working electrode. [Bard, A.J.; Faulkner, L.R. Electrochemical Methods: Fundamentals and Applications. New York: John Wiley Sons, 2nd Edition, 2000.] Structure Most… …   Wikipedia

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”