- Ecarin clotting time
Ecarin clotting time (ECT) is a laboratory test used to monitor anticoagulation during treatment with
hirudin , ananticoagulant medication which was originally isolated fromleech saliva. Ecarin, the primary reagent in this assay, is derived from the venom of the saw-scaled viper, "Echis carinatus ".In the clinical assay, a known quantity of ecarin is added to the plasma of a patient treated with hirudin. Ecarin activates
prothrombin through a specific proteolytic cleavage, which produces meizothrombin, a prothrombin-thrombin intermediate which retains the full molecular weight of prothrombin, but possesses a low level of procoagulant enzymatic activity. Crucially, this activity is inhibited by hirudin and other direct thrombin inhibitors, but not byheparin . The ECT is also unaffected by prior treatment withcoumadin or the presence of phospholipid-dependent anticoagulants, such aslupus anticoagulant . Thus, the ECT is prolonged in a specific and linear fashion with increasing concentrations of hirudin. [cite journal |author=Nowak G |title=The ecarin clotting time, a universal method to quantify direct thrombin inhibitors |journal=Pathophysiol. Haemost. Thromb. |volume=33 |issue=4 |pages=173–83 |year=2003 |pmid=15583446 |doi=10.1159/000081505] Fabrizio MC. 2001. Use of Ecarin Clotting Time (ECT) with Lepirudin Therapy in Heparin-Induced Thrombocytopenia and Cardiopulmonary Bypass. JECT 33:117–125. [http://www.amsect.org/ce/HIT/ject_2001_v33_n2_fabrizio.pdf PDF] at [http://www.amsect.org/ Journal of The American Society of ExtraCorporeal Technology] . Accessed5 June 2007 .] [http://www.specialtylabs.com/books/display.asp?id=1060 Textarin/Ecarin Time] at [http://www.specialtylabs.com/ Specialty Laboratories] . Accessed5 June 2007 .]References
Wikimedia Foundation. 2010.