Fluorescence interference contrast microscopy

Fluorescence interference contrast microscopy

Fluorescence interference contrast (FLIC) microscopy is a microscopic technique developed to achieve z-resolution on the nanometer scale.

FLIC occurs whenever fluorescent objects are in the vicinity of a reflecting surface (e.g. Si wafer). The resulting interference between the direct and the reflected light leads to a double sin2 modulation of the intensity, I, of a fluorescent object as a function of distance, h, above the reflecting surface. This allows for the "nanometer height measurements".

FLIC microscope is well suited to measuring the topography of a membrane that contains fluorescentprobes e.g. an artificial lipid bilayer, or a living cell membrane or the structure of fluorescently labeled proteins on a surface.

FLIC optical theory

General two layer system

The optical theory underlying FLIC was developed by Armin Lambacher and Peter Fromherz. They derived a relationship between the observed fluorescence intensity and the distance of the fluorophore from a reflective silicon surface.

The observed fluorescence intensity, I_{FLIC}, is the product of the excitation probability per unit time, P_{ex}, and the probability of measuring an emitted photon per unit time, P_{em}. Both probabilities are a function of the fluorophore height above the silicon surface, so the observed intensity will also be a function of the fluorophore height. The simplest arrangement to consider is a fluorophore embedded in silicon dioxide (refractive index n_{1}) a distance "d" from an interface with silicon (refractive index n_{0}). The fluorophore is excited by light of wavelength lambda_{ex} and emits light of wavelength lambda_{em}. The unit vector "e_{ex}" gives the orientation of the transition dipole of excitation of the fluorophore. P_{ex} is proportional to the squared projection of the local electric field, F_{in}, which includes the effects of interference, on the direction of the transition dipole.
P_{ex}propto mid F_{in}cdot e_{ex}mid^{2}
The local electric field, F_{in}, at the fluorophore is affected by interference between the direct incident light and the light reflecting off the silicon surface. The interference is quantified by the phase difference Phi_{in} given by
Phi_{in} = frac{4pi n_{1}dcos heta^{in}_{1{lambda_{ex
heta^{in}_{1} is the angle of the incident light with respect to the silicon plane normal. Not only does interference modulate F_{in}, but the silicon surface does not perfectly reflect the incident light. Fresnel coefficients give the change in amplitude between an incident and reflected wave. The Fresnel coefficients depend on the angles of incidence, heta_{i} and heta_{j}, the indices of refraction of the two mediums and the polarization direction. The angles heta_{i} and heta_{j} can be related by Snell's Law. The expressions for the reflection coefficients are:
r^{TE}_{ij} = frac{n_{i}cos heta_{i} - n_{j}cos heta_{j{n_{i}cos heta_{i} + n_{j}cos heta_{jquad r^{TM}_{ij} = frac{n_{j}cos heta_{i} - n_{i}cos heta_{j{n_{j}cos heta_{i} + n_{i}cos heta_{j
TE refers to the component of the electric field perpendicular to the plane of incidence and TM to the parallel component (The incident plane is defined by the plane normal and the propagation direction of the light). In cartesian coordinates, the local electric field is
F_{in} = sin gamma_{in} left [egin{array}{c}0 \1 + r^{TE}_{10} extit{e}^{ iPhi_{in \0end{array} ight] + cos gamma _{in} left [egin{array}{c}cos heta ^{in}_{1}(1-r^{TM}_{10} extit{e}^{iPhi_{in) \0 \ sin heta ^{in}_{1}(1+r^{TM}_{10} extit{e}^{iPhi_{in)end{array} ight]
gamma_{in} is the polarization angle of the incident light with respect to the plane of incidence. The orientation of the excitation dipole is a function of its angle heta_{ex} to the normal and phi_{ex} azimuthal to the plane of incidence.
extit{e}_{ex} = left [egin{array}{c}cos phi_{ex}sin heta_{ex}\sin phi_{ex}sin heta_{ex} \cos heta_{ex}end{array} ight]
The above two equations for F_{in} and extit{e}_{ex} can be combined to give the probability of exciting the fluorophore per unit time P_{ex}.
Many of the parameters used above would vary in a normal experiment. The variation in the five following parameters should be included in this theoretical description.
* The coherence of the excitation light
* The incident angle ( heta^{in}_{1}) of excitation light
* Polarization angle (gamma_{in}) of the excitation light
* The angle of transition dipole ( heta_{ex}) of the fluorophore
* The wavelength of the excitation light (lambda_{ex})The squared projection mid F_{in}cdot e_{ex}mid^{2} must be averaged over these quantities to give the probability of excitation P_{ex}. Averaging over the first 4 parameters gives
propto int sin heta_{1}^{in}d heta_{1}^{in}A_{in}( heta_{1}^{in}) imes int sin heta_{ex}d heta_{ex}O( heta_{ex})U_{ex}(lambda_{in}, heta_{1}^{in}. heta_{ex})

U_{ex} = sin^{2} heta_{ex}mid 1+r^{TE}_{10} extit{e}^{iPhi_{inmid^{2} + sin^{2} heta_{ex}cos^{2} heta^{in}_{1}mid 1-r^{TM}_{10} extit{e}^{iPhi_{inmid^{2}+2cos^{2} heta_{ex}sin^{2} heta^{in}_{1}mid 1+r^{TM}_{10} extit{e}^{iPhi_{inmid^{2}

Normalization factors are not included. O( heta_{ex}) is a distribution of the orientation angle of the fluorophore dipoles. The azimuthal angle phi_{ex} and the polarization angle gamma_{in} are integrated over analytically, so they no longer appear in the above equation. To finally obtain the probability of excitation per unit time, the above equation is integrated over the spread in excitation wavelength, accounting for the intensity I(lambda_{ex}) and the extinction coefficient of the fluorophore epsilon(lambda_{ex}).
P_{ex}propto int dlambda_{ex}I(lambda_{ex})epsilon(lambda_{ex})
The steps to calculate P_{em} are equivalent to those above in calculating P_{ex} except that the parameter labels "em" are replaced with "ex" and "in" is replaced with "out". P_{em}propto int dlambda_{em}Phi_{det}(lambda_{em}) extit{f}(lambda_{em})
The resulting fluorescence intensity measured is proportional to the product of the excitation probability and emission probability

I_{FLIC} propto P_{ex}P_{em}
It is important to note that this theory determines a proportionality relation between the measured fluorescence intensity I_{FLIC} and the distance of the fluorophore above the reflective surface. The fact that it is not an equality relation will have a significant effect on the experimental procedure.

Experimental Setup

A silicon wafer is typically used as the reflective surface in a FLIC experiment. An oxide layer is then thermally grown on top of the silicon wafer to act as a spacer. On top of the oxide is placed the fluorescently labeled specimen, such as a lipid membrane, a cell or membrane bound proteins. With the sample system built, all that is needed is an epifluorescence microscope and a CCD camera to make quantitative intensity measurements.

The silicon dioxide thickness is very important in making accurate FLIC measurements. As mentioned before, the theoretical model describes the "relative" fluorescence intensity measured versus the fluorophore height. The fluorophore position cannot be simply read off of a single measured FLIC curve. The basic procedure is to manufacture the oxide layer with at least two known thicknesses (the layer can be made with photolithographic techniques and the thickness measured by ellipsometry). The thicknesses used depends on the sample being measured. For a sample with fluorophore height in the range of 10nm, oxide thickness around 50nm would be best because the FLIC intensity curve is steepest here and would produce the greatest contrast between fluorophore heights. Oxide thickness above a few hundred nanometers could be problematic because the curve begins to get smeared out by polychromatic light and a range of incident angles. A ratio of measured fluorescence intensities at different oxide thicknesses is compared to the predicted ratio to calculate the fluorophore height above the oxide (d_{ extit{f,).
frac{I_{theory}(d_{1})}{I_{theory}(d_{0})}=frac{I_{exp}(d_{1}+d_{ extit{f)}{I_{exp}(d_{0}+d_{ extit{f)}
The above equation can then be solved numerically to find d_{ extit{f.Imperfections of the experiment , such as imperfect reflection, nonnormal incidence of light and polychromatic light tend to smear out the sharp fluorescence curves. The spread in incidence angle can be controlled by the numerical aperture (N.A.). However, depending on the numerical aperture used, the experiment will yield good lateral resolution (x-y) or good vertical resolution (z), but not both. A high N.A. (~1.0) gives good lateral resolution which is best if the goal is to determine long range topography. Low N.A. (~0.001), on the other hand, provides accurate z-height measurement to determine the height of a fluorescently labeled molecule in a system.

Analysis

The basic analysis involves fitting the intensity data with the theoretical model allowing the distance of the fluorophore above the oxide surface (d_{ extit{f) to be a free parameter.The FLIC curves shift to the left as the distance of the fluorophore above the oxide increases. d_{ extit{f is usually the parameter of interest, but several other free parameters are often included to optimize the fit. Normally an amplitude factor (a) and a constant additive term for the background (b) are included. The amplitude factor scales the relative model intensity and the constant background shifts the curve up or down to account for fluorescence coming from out of focus areas, such as the top side of a cell. Occasionally the numerical aperture (N.A.) of the microscope is allowed to be a free parameter in the fitting. The other parameters entering the optical theory, such as different indices of refraction, layer thicknesses and light wavelengths, are assumed constant with some uncertainty.A FLIC chip may be made with oxide terraces of 9 or 16 different heights arranged in blocks. After a fluorescence image is captured, each 9 or 16 terrace block yields a separate FLIC curve that defines a unique d_{ extit{f. The average d_{ extit{f is found by compiling all the d_{ extit{f values into a histogram.
The statistical error in the calculation of d_{ extit{f comes from two sources: the error in fitting of the optical theory to the data and the uncertainty in the thickness of the oxide layer. Systematic error comes from three sources: the measurement of the the oxide thickness (usually by ellipsometer), the fluorescence intensity measurement with the CCD, and the uncertainty in the parameters used in the optical theory. The systematic error has been estimated to be sim 1 nm.

References

*Ajo-Franklin C., Yoshina-Ishii C., and Boxer S. "Langmuir" 21, 4976-4983 (2005).
*Braun D. and Fromherz P. "App. Phys. A." 65, 341-348 (1997).
*Braun D. and Fromherz P. "Phys. Rev. Lett." 81, 5241-5244 (1998).
*Crane J., Kiessling V., and Tamm L. "Langmuir" 21, 1377-1388 (2005).
*Kaizuka Y. and Groves J. "Phys. Rev. Lett." 96, 118101 (2006).
*Kiessling V. and Tamm L. "Biophy. J." 84, 408-418 (2003).
*Lambacher A. and Fromherz P. "App. Phys. A" 63, 207-216 (1996).
*Lambacher A. and Fromherz P. "J. Opt. Soc. Am. B." 19, 1435-1453 (2002).
*Parthasarathy R. and Groves J. "Cell Biochem. and Biophy." 41, 391-414 (2004).


Wikimedia Foundation. 2010.

Игры ⚽ Нужен реферат?

Look at other dictionaries:

  • Differential interference contrast microscopy — Micrasterias furcata imaged in transmitted DIC microscopy. Differential interference contrast microscopy (DIC), also known as Nomarski Interference Contrast (NIC) or Nomarski microscopy, is an optical microscopy illumination technique used to… …   Wikipedia

  • Interference microscopy — involving measurements of differences in the path between two beams of light that have been split.[1] Types include: Classical interference microscopy Differential interference contrast microscopy Fluorescence interference contrast microscopy See …   Wikipedia

  • Interference microscope — might refer to: * Classical interference microscopy (interference between well separated beams) * Differential interference contrast microscopy (also called Nomarski Interference Contrast or Nomarski microscopy) * Fluorescence interference… …   Wikipedia

  • Microscopy — is the technical field of using microscopes to view samples and objects that cannot be seen with the unaided eye (objects that are not within the resolution range of the normal eye). There are three well known branches of microscopy, optical,… …   Wikipedia

  • Phase contrast microscopy — Phase contrast image of a cheek epithelial cell Phase contrast microscopy is an optical microscopy illumination technique of great importance to biologists in which small (invisible to the human eye) phase shifts in the light passing through a… …   Wikipedia

  • Fluorescence microscope — An upright fluorescence microscope (Olympus BX61) with the fluorescent filter cube turret above the objective lenses, coupled with a digital camera. A fluorescence microscope is an optical microscope used to study properties of organic or… …   Wikipedia

  • Microscopy — The examination of minute objects by means of a microscope, an instrument which provides an enlarged image of an object not visible with the naked eye. Aside from the usual microscopy, there are various special types of microscopy including, for… …   Medical dictionary

  • Confocal microscopy — Diagnostics MeSH D018613 OPS 301 code 3 301 …   Wikipedia

  • Dark field microscopy — (dark ground microscopy) describes microscopy methods, in both light and electron microscopy, which exclude the unscattered beam from the image. As a result, the field around the specimen (i.e. where there is no specimen to scatter the beam) is… …   Wikipedia

  • Two-photon excitation microscopy — is a fluorescence imaging technique that allows imaging of living tissue up to a very high depth, that is up to about one millimeter. Being a special variant of the multiphoton fluorescence microscope, it uses red shifted excitation light which… …   Wikipedia