Nephelometer

Nephelometer
A nephelometer at the Kosan, Cheju Island, South Korea NOAA facility

A nephelometer [1] is a stationary or portable instrument for measuring suspended particulates in a liquid or gas colloid. A nephelometer measures suspended particulates by employing a light beam (source beam) and a light detector set to one side (often 90°) of the source beam. Particle density is then a function of the light reflected into the detector from the particles. To some extent, how much light reflects for a given density of particles is dependent upon properties of the particles such as their shape, color, and reflectivity. Nephelometers are calibrated to a known particulate, commonly Arizona road dust then use environmental factors k-factors to compensate lighter or darker colored dusts accordingly. K-factor is determined by the user by running the nephelometer next to an air sampling pump and comparing results.

Contents

Nephelometer uses

The main uses of nephelometers relate to air quality measurement for pollution monitoring, climate monitoring, and visibility. Airborne particles are commonly either biological contaminants, particulate contaminants, gaseous contaminants, or dust.

Particulate Contaminants

This chart shows the types and sizes of various particulate contaminants. This information is helpful toward understanding the character of particulate pollution inside a building or in the ambient air. It is also useful for understanding the cleanliness level in a controlled environment.

Biological contaminants include mold, fungus, bacteria, viruses, animal dander, dust mites, pollen, human skin cells, cockroach parts, or anything alive or living at one time. These are the biggest enemy of Indoor Air Quality specialist because they are contaminants that cause health problems. Levels of biological contamination depend on humidity and temperature that supports the livelihood of micro-organisms. The presence of pets, plants, rodents, and insects will raise the level of biological contamination.

Sheath air

Sheath air is clean filtered air that surrounds the aerosol stream to prevent particulates from circulating or depositing within the optic chamber. Sheath air prevents contamination caused by build-up and deposits, improves response time by containing the sample, and improves maintenance by keeping the optic chamber clean. The nephelometer creates the sheath air by passing air through a zero filter before beginning the sample.

Global radiation balance

Nephelometers are also used in global warming studies, specifically measuring the global radiation balance. Three wavelength nephelometers fitted with a backscatter shutter can determine the amount of solar radiation that is reflected back into space through dust and particulate matter. This reflected light influences the amount of radiation reaching the earths lower atmosphere and warming the planet. A simple demonstration of this is found in www.aurora-nephelometer.com as seen below

Radiation Balance (thickness of Earth's atmosphere is greatly exaggerated)

Visibility

Nephelometers are also used for measurement of visibility with simple one-wavelength nephelometers used throughout the world by many EPAs. Nephelometers, through the measurement of light scattering, can determine visibility in distance through the application of a conversion factor called Koschmieder’s formula.

Turbidity units

A nephelometer installation at Acadia National Park
  • Because optical properties depend on suspended particle size, a stable synthetic material called "Formazin" with uniform particle size is often used as a standard for calibration and reproducibility.[2] The unit is called Formazin Turbidity Unit (FTU).
  • Nephelometric Turbidity Units (NTU) specified by United States Environmental Protection Agency is a special case of FTU, where a white light source and certain geometrical properties of the measurement apparatus are specified. (Sometimes the alternate form "nephelos turbidity units" is used[3][4])
  • Formazin Nephelometric Units (FNU), prescribed for 9 measurements of turbidity in water treatment by ISO 7027, another special case of FTU with near infrared light (NIR) and 90º scatter.
  • Formazin Attenuation Units (FAU) specified by ISO 7027 for water treatment standards for turbidity measurements at 0°, also a special case of FTU.
  • Formazin Backscatter Units (FBU), not part of a standard, is the unit of optical backscatter detectors (OBS), measured at ca 180º, also a special case of FTU.
  • European Brewery Convention (EBC) turbidity units
  • Concentration Units (C.U.)
  • Optical Density (O.D.)
  • Jackson "Candle" Turbidity Units (JTU; an early measure)
  • Helms Units
  • American Society of Brewing Chemists (ASBC-FTU) turbidity units
  • Parts Per Million of standard substance, such as PPM/DE (Kieselguhr)
  • "Trübungseinheit/Formazin" (TE/F) a German standard, now replaced by the FNU unit.
  • diatomaceous earth ("ppm SiO2") an older standard, now obsolete

A more popular term for this instrument in water quality testing is a turbidimeter. However, there can be differences between models of turbidimeters, depending upon the arrangement (geometry) of the source beam and the detector. A nephelometric turbidimeter always monitors light reflected off the particles and not attenuation due to cloudiness. In the United States environmental monitoring the turbidity standard unit is called Nephelometric Turbidity Units (NTU), while the international standard unit is called Formazin Nephelometric Unit (FNU). The most generally applicable unit is Formazin Turbidity Unit (FTU), although different measurement methods can give quite different values as reported in FTU (see below).

Gas-phase nephelometers are also used to study the atmosphere. These can provide information on visibility and atmospheric albedo. Gas-phase nephelometers are also used in the detection of smoke & other particles of combustion. In such use, the apparatus is referred to as an aspirated smoke detector. These have the capability to detect extremely low particle concentrations (to 0.005%) and are therefore highly suitable to protecting sensitive or valuable electronic equipment, such as mainframe computers and telephone switches.

See also

Notes

  1. ^ Nephelometer comes from the Greek word for cloud, nephos, cf. "nepheloid layer".
  2. ^ Formazin was first used for standardizing turbidity measurements in 1926.
  3. ^ Reducing turbidity in chromic acid solutions, Zeller, III, Robert L.; Morgan, Russell J.; Rabbe, Gilbert D.; Fiscus, Donna R.; Wilkes, Jr., Richard L.; United States Patent 5034211, Filing date 1990-10-29, Publication date 1991-07-23
  4. ^ Florida Department of Agriculture and Consumer Services 2006 Codebook Chapter 5L-1: The Comprehensive Shellfish Control code



Wikimedia Foundation. 2010.

Игры ⚽ Поможем написать реферат

Look at other dictionaries:

  • Nephelometer — Ein Nephelometer ist ein Messgerät zur Bestimmung der Suspension von kleinen Partikeln in einer Flüssigkeit oder einem Gaskolloid.[1] Es besteht aus einer Lichtquelle und einem Fotosensor …   Deutsch Wikipedia

  • Nephelometer — Neph e*lom e*ter (n[e^]f [ e]*l[o^]m [ e]*t[ e]r), n. [Gr. nefe lh a cloud + meter.] 1. An instrument for measuring or registering the amount of cloudiness. [1913 Webster] 2. (Chem., Microbiol.) An instrument which measures the degree to which… …   The Collaborative International Dictionary of English

  • nephelometer — [nef΄ə läm′ət ər] n. [ NEPHELO + METER] an apparatus for measuring the concentration of a suspension, as of bacteria or other substance, by its scattering of a beam of light …   English World dictionary

  • Nephelometer — nefelometras statusas T sritis Standartizacija ir metrologija apibrėžtis Išsklaidytosios šviesos matavimo įtaisas su integruojančiąja sfera. atitikmenys: angl. hazemeter; nephelometer vok. Hazemeter, n; Nephelometer, n; Trübungsmesser, m rus.… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • nephelometer — nefelometras statusas T sritis Standartizacija ir metrologija apibrėžtis Išsklaidytosios šviesos matavimo įtaisas su integruojančiąja sfera. atitikmenys: angl. hazemeter; nephelometer vok. Hazemeter, n; Nephelometer, n; Trübungsmesser, m rus.… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • Nephelometer — nefelometras statusas T sritis Standartizacija ir metrologija apibrėžtis Optinis dispersinių sistemų koncentracijos ir dispersiškumo matuoklis, kurio veikimas pagrįstas išsklaidytosios šviesos intensyvumo matavimu. atitikmenys: angl. hazemeter;… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • nephelometer — nefelometras statusas T sritis Standartizacija ir metrologija apibrėžtis Optinis dispersinių sistemų koncentracijos ir dispersiškumo matuoklis, kurio veikimas pagrįstas išsklaidytosios šviesos intensyvumo matavimu. atitikmenys: angl. hazemeter;… …   Penkiakalbis aiškinamasis metrologijos terminų žodynas

  • Nephelometer — nefelometras statusas T sritis fizika atitikmenys: angl. nephelometer vok. Nephelometer, n; Trübungsmesser, m rus. нефелометр, m pranc. néphélomètre, m …   Fizikos terminų žodynas

  • nephelometer — nefelometras statusas T sritis fizika atitikmenys: angl. nephelometer vok. Nephelometer, n; Trübungsmesser, m rus. нефелометр, m pranc. néphélomètre, m …   Fizikos terminų žodynas

  • nephelometer — noun Etymology: Greek nephelē cloud + International Scientific Vocabulary meter Date: 1884 1. an instrument for measuring the extent or degree of cloudiness 2. an instrument for determining the concentration or particle size of suspensions by… …   New Collegiate Dictionary

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”