Frequency-resolved optical gating

Frequency-resolved optical gating

In optics, frequency-resolved optical gating (FROG) is a derivative of autocorrelation, but is far superior in its ability to measure ultrafast optical pulse shapes. Further, it can determine the phase of the pulse.In the most common configuration, FROG is simply a background-free autocorrelator followed by a spectrometer.It is the two-dimensional nature of the FROG trace that allows the extraction of the actual pulse shape "and phase" from the data.

The basics

Frequency-Resolved refers to the fact that the final signal is a spectrum. Before explaining Optical Gating, it helps to recognize that the pulse is really interacting with itself. In most configurations, the pulse is split and recombined, as in an interferometer. However, in this case, the recombination does not occur on a beam-splitter, but rather in a nonlinear medium, which allows the two beams to interact with each other. It is this interaction that allows the pulses to "gate" the spectral information of the other pulse. So Optical Gating refers to the fact that the measured spectrum is really from a time-slice of the pulse, and that time-slice is determined by the pulse nonlinear interaction. The gate function depends on the type of nonlinear interaction allowed. Mathematically the FROG trace is simply a spectrogram but with an unknown gate function:

: I_{sig}(omega, au) = left | int_{-infty}^{infty} P(t) G(t- au) e^{-i omega t} dt ight | ^2

where P(t) is the "probe" pulse and G(t) is the "gate" pulse.The probe and gate pulses are determined by the nonlinear interaction used, andit is the form of the probe and gate that distinguishes the different types of FROG from each other.Some of the most common are:



thumb|right|300px|A schematic of a typical experimental, multishot SHG FROG setup.

For example second-harmonic generation FROG (SHG FROG) would be:

: I^{SHG}_{sig}(omega, au) = left | int_{-infty}^{infty} E(t)E(t- au)e^{-i omega t} dt ight | ^2

and PG FROG would be:

: I^{PG}_{sig}(omega, au) = left | int_{-infty}^{infty} E(t)|E(t- au)|^2 e^{-i omega t} dt ight | ^2

Traditional inversion algorithms for spectrograms requires perfect knowledge of the gate function (G(t)), however, FROG does not have this luxury. Instead an iterative alogrithm is used. The algorithm uses both the data (FROG trace) and form of the nonlinearity to achieve a best match between the real FROG trace and the "retrieved" FROG trace.The retrieved FROG trace is created synthetically from the best guess for E(t).

FROG algorithm

The FROG algorithm is all about phase retrieval. The FROG trace measured in the lab is the exact intensity of I_{sig}(omega, au); however, it is missing the phase information.To start off with define a signal field:

: E_{sig}left(t, au ight) = Pleft(t ight )Gleft(t- au ight)

and further defining:

: ilde{E}_{sig}(omega, au) = int_{-infty}^{infty} E_{sig}(t, au)e^{-i omega t} dt

given this the FROG trace becomes:

: I_{sig}(omega, au) = left | ilde{E}_{sig}(omega, au) ight | ^2

inverting this:

: left | ilde{E}_{sig}(omega, au) ight | = sqrt{I_{sig}(omega, au)}

so the amplitude of ilde{E}_{sig}(omega, au) is known, but not its phase.If this phase is found, then the pulse (E(t)) can be retrieved.

An iterative algorithm is used to determine this unknown phase.

# Start with an initial guess for E(t),.
# Synthetically create E_{sig}(t, au), from E(t),.
# Fourier transform the time axis to the frequency domain, yielding ilde{E}_{sig}(omega, au).
# Replace the amplitude (preserving the phase) of ilde{E}_{sig}(omega, au) with the amplitude measured in the lab (I_{sig}(omega, au),). Call this ilde{E'}_{sig}(omega, au).
# Take ilde{E'}_{sig}(omega, au) and inverse Fourier transform it back into the time domain (E'_{sig}(t, au),).
# Extract the best E(t), from E'_{sig}(t, au),.
# Synthetically create E_{sig}(t, au), from E(t),.
# Fourier transform the time axis to the frequency domain, yielding ilde{E}_{sig}(omega, au).
# Compare left | ilde{E}_{sig}(omega, au) ight | to I_{sig}(omega, au), (generally the rms difference). If this error (termed the "G" error) is sufficiently small, exit the loop.
# GOTO 4.

ee also

FROG techniques

* Grating-eliminated no-nonsense observation of ultrafast incident laser light e-fields (GRENOUILLE), a simplified version of FROG

Competing techniques

* Spectral phase interferometry for direct electric-field reconstruction (SPIDER)
* Streak camera — not a significant competitor. Streak cameras have picosecond response times.
* Multiphoton intrapulse interference phase scan (MIIPS), a method to characterize and manipulate the ultrashort pulse.
* Frequency-resolved electro-absorption gating (FREAG)

References

Further reading

*cite book|title=Frequency-Resolved Optical Gating: The Measurement of Ultrashort Laser Pulses|id=ISBN 1-4020-7066-7|author=Rick Trebino|publisher=Springer|year= 2002

External links

* [http://www.physics.gatech.edu/gcuo/subIndex.html FROG Page by Rick Trebino] (co-inventor of FROG)


Wikimedia Foundation. 2010.

Игры ⚽ Нужен реферат?

Look at other dictionaries:

  • Frequency-resolved optical gating — Frequency Resolved Optical Gating, plus couramment désigné par son acronyme FROG, est une méthode de mesure des impulsions lumineuses ultra brèves (de l ordre de la femtoseconde) par auto corrélation. Son principe repose sur un montage de type… …   Wikipédia en Français

  • Frequency-Resolved Optical Gating — Frequency Resolved Optical Gating, plus couramment désigné par son acronyme FROG, est une méthode de mesure des impulsions lumineuses ultra brèves (de l ordre de la femtoseconde) par auto corrélation. Son principe repose sur un montage de type… …   Wikipédia en Français

  • Optical coherence tomography — Intervention Optical Coherence Tomography (OCT) image of a sarcoma MeSH …   Wikipedia

  • GRENOUILLE — Grating eliminated no nonsense observation of ultrafast incident laser light e fields (GRENOUILLE) is an ultrashort pulse measurement technique based on frequency resolved optical gating (FROG). The somewhat frivolous acronym was chosen because… …   Wikipedia

  • Spectral phase interferometry for direct electric-field reconstruction — In ultrafast optics, spectral phase interferometry for direct electric field reconstruction (SPIDER) is an ultrashort pulse measurement technique.The basicsSPIDER is an interferometric ultrashort pulse measurement technique in the frequency… …   Wikipedia

  • Multiphoton intrapulse interference phase scan — (MIIPS) is a method used in ultrashort laser technology that simultaneously measures (phase characterization), and compensates (phase correction) femtosecond laser pulses using an adaptive pulse shaper. Current ultrashort laser pulse… …   Wikipedia

  • Ultrashort pulse — In optics, an ultrashort pulse of light is an electromagnetic pulse whose time duration is on the order of the femtosecond (10 − 15 second). Such pulses have a broadband optical spectrum, and can be created by mode locked oscillators. They are… …   Wikipedia

  • Laser acronyms — Here, is a list of acronyms used in laser physics, applications and technology.A*AOM – acousto optic modulator *APD – avalanche photodiode *APM – additive pulse mode locking *ASE – amplified spontaneous emission *AWG – arrayed waveguide… …   Wikipedia

  • Modelocking — Mode locking is a technique in optics by which a laser can be made to produce pulses of light of extremely short duration, on the order of picoseconds (10 12s) or femtoseconds (10 15s).The basis of the technique is to induce a fixed phase… …   Wikipedia

  • Mode-locking — is a technique in optics by which a laser can be made to produce pulses of light of extremely short duration, on the order of picoseconds (10−12 s) or femtoseconds (10−15 s). The basis of the technique is to induce a fixed phase… …   Wikipedia

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”