- Hemodialysis
In
medicine , hemodialysis (also haemodialysis) is a method for removing waste products such aspotassium andurea , as well as free water from theblood when thekidney s are inrenal failure . Hemodialysis is one of three renal replacement therapies (the other two beingrenal transplant ;peritoneal dialysis ).Hemodialysis can be an
outpatient orinpatient therapy. Routine hemodialysis is conducted in a dialysis outpatient facility, either a purpose built room in ahospital or a dedicated, stand alone clinic. Less frequently hemodialysis is done at home. Dialysis treatments in a clinic are initiated and managed by specialized staff made up of nurses and technicians; dialysis treatments at home can be self initiated and managed or done jointly with the assistance of a trained helper who is usually a family member. [National Kidney and Urologic Diseases Information Clearinghouse guidance " [http://kidney.niddk.nih.gov/Kudiseases/pubs/choosingtreatment/index.htm Kidney Failure: Choosing a Treatment That’s Right for You] "]Principle
The principle of hemodialysis is the same as other methods of
dialysis ; it involvesdiffusion of solutes across a semipermeable membrane. Hemodialysis utilizes counter current flow, where the dialysate is flowing in the opposite direction toblood flow in theextracorporeal circuit. Counter-current flow maintains the concentration gradient across the membrane at a maximum and increases the efficiency of the dialysis.Fluid removal (ultrafiltration) is achieved by altering the
hydrostatic pressure of the dialysate compartment, causing free water and some dissolved solutes to move across the membrane along a created pressure gradient.The dialysis solution that is used is a sterilized solution of mineral ions.
Urea and other waste products, and also,potassium andphosphate , diffuse into the dialysis solution. However, concentrations ofsodium and chloride are similar to those of normal plasma to prevent loss. Bicarbonate is added in a higher concentration than plasma to correct blood acidity. A small amount of glucose is also commonly used.Note that this is a different process to the related technique of
hemofiltration .History
Many have played a role in developing dialysis as a practical treatment for renal failure, starting with Thomas Graham of Glasgow, who first presented the principles of solute transport across a semipermeable membrane in 1854. [Graham T. [http://links.jstor.org/sici?sici=0261-0523(1854)144%3C177%3ATBLOOF%3E2.0.CO%3B2-E The Bakerian lecture: on osmotic force] . Philosophical Transactions of the Royal Society in London. 1854;144:177–228. The artificial kidney was first developed by Abel, Rountree and Turner in 1913, [Abel, J. J., Rountree, L. G., and Turner, B. B. [http://books.google.com/books?id=KMcCAAAAYAAJ&dq=&pg=PA51&ots=UM7CVprPEW&sig=Xpnf-kEJTYO7iFSxhdSoC2Ujh3Y&prev The removal of diffusible substances from the circulating blood by means of dialysis] . Tn. Assoc. Am. Phys., 28:51, 1913.] , the first hemodialysis in a human being was by Hass (February 28, 1924) [ [http://www.uniklinikum-giessen.de/med3/history/haas/2001-Dial-Transpl.pdf Georg Haas (1886–1971): The Forgotten Hemodialysis Pioneer (PDF)] ] and the artificial kidney was developed into a clinically useful apparatus by Kolff in 1943 - 1945. [Kolff, W. J., and Berk, H. T. J. [http://jasn.asnjournals.org/cgi/reprint/8/12/1959 Artificial kidney, dialyzer with great area] . Geneesk. gids., 21:1944.] This research showed that life could be prolonged in patients dying of
renal failure .Dr.
Willem Kolff was the first to construct a working dialyzer in1943 . The first successfully treated patient was a 67-year-old woman in uremic coma who regained consciousness after 11 hours of hemodialysis with Kolff’s dialyzer in1945 . At the time of its creation, Kolff’s goal was to provide life support during recovery from acute renal failure. AfterWorld War II ended, Kolff donated the five dialyzers he’d made to hospitals around the world, includingMount Sinai Hospital, New York . Kolff gave a set of blueprints for his hemodialysis machine to George Thorn at thePeter Bent Brigham Hospital inBoston . This led to the manufacture of the next generation of Kolff’s dialyzer, astainless steel Kolff-Brigham dialysis machine.By the 1950s, Willem Kolff’s invention of the dialyzer was used for acute renal failure, but it was not seen as a viable treatment for patients with stage 5 chronic kidney disease (CKD). At the time, doctors believed it was impossible for patients to have dialysis indefinitely for two reasons. Firstly, they thought no man-made device could replace the function of kidneys over the long term. In addition, a patient undergoing dialysis suffered from damaged veins and arteries, so that after several treatments, it became difficult to find a vessel to access the patient’s blood.
Dr.
Nils Alwall : The original Kolff kidney was not very useful clinically, because it did not allow for removal of excess fluid. Dr. Nils Alwall [ [http://www.med.lu.se/english/about_the_faculty/faculty_milestones/nils_alwall University of Lund website: Nils Alwall.] ] encased a modified version of this kidney inside a stainless steel canister, to which a negative pressure could be applied, in this way effecting the first truly practical application of hemodialysis, which was done in 1946 at the University of Lund. Alwall also was arguably the inventor of the arteriovenous shunt for dialysis. He reported this first in 1948 where he used such an arteriovenous shunt in rabbits. Subsequently he used such shunts, made of glass, as well as his canister-enclosed dialyzer, to treat 1500 patients in renal failure between 1946 and 1960, as reported to the First International Congress of Nephrology held in Evian in September 1960. Alwall was appointed to a newly-created Chair of Nephrology at the University of Lund in 1957. Subsequently, he collaborated with Swedish businessman Holger Crafoord to found one of the key companies that would manufacture dialysis equipment in the past 40 years, Gambro, Inc. The early history of dialysis has been reviewed by Stanley Shaldon [Shaldon S. Development of Hemodialysis, From Access to Machine (presentation given during a symposium entitled: Excellence in Dialysis: Update in Nephrology; Karachi, Pakistan. October, 2002, [http://www.hdcn.com/symp/02pak/02pak.htm as archived on HDCN] ] .Dr.
Belding H. Scribner working with a surgeon, Dr. Wayne Quinton, modified the glass shunts used by Alwall by making them from Teflon. Another key improvement was to connect them to a short piece of silicone elastomer tubing. This formed the basis of the so-calledScribner shunt , perhaps more properly called the Quinton-Scribner shunt. After treatment, the circulatory access would be kept open by connecting the two tubes outside the body using a small U-shaped Teflon tube, which would shunt the blood from the tube in the artery back to the tube in the vein [ [http://kidney.niddk.nih.gov/about/Research_Updates/win00-01/contrib.htm NIDDK Contributions to Dialysis] ] .In 1962, Scribner started the world’s first outpatient dialysis facility, the Seattle Artificial Kidney Center, later renamed the
Northwest Kidney Centers . Immediately the problem arose of who should be given dialysis, since demand far exceeded the capacity of the six dialysis machines at the center. Scribner decided that the decision about who would receive dialysis and who wouldn’t, would not be made by him. Instead, the choices would be made by an anonymous committee, which could be viewed as one of the firstbioethics committees.For a detailed history of successful and unsuccessful attempts at dialysis, including pioneers such as Abel and Roundtree, Haas, and Necheles, see this review by Kjellstrand [Kjellstrand CM. History of Dialysis, Men and Ideas. Talk given to the Nordic Nephrology Days Symposium, Lund, 1997, [http://www.hdcn.com/symp/lund/kjel.htm as archived on HDCN.] ] .
Prescription
A prescription for dialysis by a
nephrologist (a medical kidney specialist) will specify various parameters for a dialysis treatment. These include frequency (how many treatments per week), length of each treatment, and the blood and dialysis solution flow rates, as well as the size of the dialyzer. The composition of the dialysis solution is also sometimes adjusted in terms of its sodium and potassium and bicarbonate levels. In general, the larger the body size of an individual, the more dialysis he will need. In theNorth America and UK, 3-4 hour treatments (sometimes up to 5 hours for larger patients) given 3 times a week are typical. Twice-a-week sessions are limited to patients who have a substantial residual kidney function. Four sessions per week are often prescribed for larger patients, as well as patients who have trouble with fluid overload. Finally, there is growing interest inshort daily home hemodialysis , which is 1.5 - 4 hr sessions given 5-7 times per week, usually at home. There also is interest innocturnal dialysis , which involves dialyzing a patient, usually at home, for 8-10 hours per night, 3-6 nights per week. Nocturnal in-center dialysis, 3-4 times per week is also offered at a handful of dialysis units in theUnited States .ide-effects and complications
Hemodialysis often involves fluid removal (through ultrafiltration), because most patients with
renal failure pass little or no urine. Side effects caused by removing too much fluid and/or removing fluid too rapidly include low blood pressure, fatigue, chest pains, leg-cramps,nausea andheadaches . These symptoms can occur during the treatment and can persist post treatment; they are sometimes collectively referred to as the dialysis hangover or dialysis washout. The severity of these symptoms are usually proportionate to the amount and speed of fluid removal. However, the impact of a given amount or rate of fluid removal can vary greatly from person to person and day to day. These side effects can be avoided and/or their severity lessened by limiting fluid intake between treatments or increasing the dose of dialysis e.g. dialyzing more often or longer per treatment than the standard three times a week, 3-4 hours per treatment schedule.Since hemodialysis requires access to the circulatory system, patients undergoing hemodialysis may expose their circulatory system to
microbes , which can lead tosepsis , an infection affecting the heart valves (endocarditis ) or an infection affecting the bones (osteomyelitis ). The risk of infection varies depending on the type of access used (see below). Bleeding may also occur, again the risk varies depending on the type of access used. Infections can be minimized by strictly adhering toinfection control best practices.Heparin is the most commonly used anticoagulant in hemodialysis, as it is generally well tolerated and can be quickly reversed withprotamine sulfate . Heparin allergy can infrequently be a problem and can cause a low platelet count. In such patients, alternative anticoagulants can be used. In patients at high risk of bleeding, dialysis can be done without anticoagulation.First Use Syndrome is a rare but severe anaphylactic reaction to the
artificial kidney . Its symptoms include sneezing, wheezing, shortness of breath, back pain, chest pain, or sudden death. It can be caused by residual sterilant in the artificial kidney or the material of the membrane itself. In recent years, the incidence of First Use Syndrome has decreased, due to an increased use ofgamma irradiation , steam sterilization, or electron-beam radiation instead of chemical sterilants, and the development of new semipermeable membranes of higherbiocompatibility . New methods of processing previously acceptable components of dialysis must always been considered. For example, in 2008, a series of first-use type or reactions, including deaths occurred due to heparin contaminated during the manufacturing process with oversulfatedchondroitin sulfate . [cite journal |author=Kishimoto TK, Viswanathan K, Ganguly T et al. |title=Contaminated heparin associated with adverse clinical events and activation of the contact system |journal=N Engl J Med |volume=358 |issue=23 |pages=2457–67 |year=2008 |pmid=18434646 |doi=]Longterm complications of hemodialysis include
amyloidosis ,neuropathy and various forms ofheart disease . Increasing the frequency and length of treatments have been shown to improve fluid overload and enlargement of the heart that is commonly seen in such patients. [ [http://www.homedialysis.org/pros/abstracts/20051014/ Effects of short daily versus conventional hemodialysis on left ventricular hypertrophy and inflammatory markers: a prospective, controlled study] ] [cite journal |author=Weinreich T, De los Ríos T, Gauly A, Passlick-Deetjen J |title=Effects of an increase in time vs. frequency on cardiovascular parametersin chronic hemodialysis patients |journal=Clin. Nephrol. |volume=6 |issue=6 |pages=433–9 |year=2006 |pmid=17176915 |doi=]Listed below are specific complications associated with different types of hemodialysis access.
Access
In hemodialysis, three primary methods are used to gain access to the blood: an intravenous catheter, an arteriovenous (AV) fistula, or a synthetic graft. The type of access is influenced by factors such as the expected time course of a patient's renal failure and the condition of his or her vasculature. Patients may have multiple accesses, usually because an AV fistula or graft is maturing and a catheter is still being used.
Catheter
Catheter access, sometimes called a CVC (Central Venous Catheter), consists of a plastic catheter with two lumens (or occasionally two separate catheters) which is inserted into a large vein (usually the
vena cava , via theinternal jugular vein or thefemoral vein ) to allow large flows of blood to be withdrawn from one lumen, to enter the dialysis circuit, and to be returned via the other lumen. However, blood flow is almost always less than that of a well functioning fistula or graft.Catheters are usually found in two general varieties, tunnelled and non-tunnelled.
Non-tunnelled catheter access is for short-term access (up to about 10 days, but often for one dialysis session only), and the catheter emerges from the skin at the site of entry into the vein.
Tunnelled catheter access involves a longer catheter, which is tunnelled under the skin from the point of insertion in the vein to an exit site some distance away. It is usually placed in the internal jugular vein in the neck and the exit site is usually on the chest wall. The tunnel acts as a barrier to invading microbes, and as such, tunnelled catheters are designed for short- to medium-term access (weeks to months only), because infection is still a frequent problem. Aside from infection, venous
stenosis is another serious problem with catheter access. The catheter is a foreign body in the vein and often provokes an inflammatory reaction in the vein wall. This results inscarring and narrowing of the vein, often to the point of occlusion. This can cause problems with severe venous congestion in the area drained by the vein and may also render the vein, and the veins drained by it, useless for creating a fistula or graft at a later date. Patients on long-term hemodialysis can literally 'run out' of access, so this can be a fatal problem. Catheter access is usually used for rapid access for immediate dialysis, for tunnelled access in patients who are deemed likely to recover fromacute renal failure , and for patients withend-stage renal failure who are either waiting for alternative access to mature or who are unable to have alternative access. Catheter access is often popular with patients, because attachment to the dialysis machine doesn't require needles. However, the serious risks of catheter access noted above mean that such access should be contemplated only as a long-term solution in the most desperate access situation.
=AV fistula= AV (arteriovenous) fistulas are recognized as the preferred access method. To create afistula , avascular surgeon joins anartery and avein together throughanastomosis . Since this bypasses thecapillaries , blood flows rapidly through the fistula. One can feel this by placing one's finger over a mature fistula. This is called feeling for "thrill" and produces a distinct 'buzzing' feeling over the fistula. Fistulas are usually created in the nondominant arm and may be situated on the hand (the 'snuffbox ' fistula'), the forearm (usually a radiocephalic fistula, or so-called Brescia-Cimino fistula , in which theradial artery is anastomosed to thecephalic vein ), or the elbow (usually a brachiocephalic fistula, where thebrachial artery is anastomosed to thecephalic vein ). A fistula will take a number of weeks to mature, on average perhaps 4-6 weeks. During treatment, two needles are inserted into the fistula, one to draw blood and one to return it.The advantages of the AV fistula use are lower infection rates, because no foreign material is involved in their formation, higher blood flow rates (which translates to more effective dialysis), and a lower incidence of
thrombosis . The complications are few, but if a fistula has a very high blood flow and the vasculature that supplies the rest of the limb is poor, a steal syndrome can occur, where blood entering the limb is drawn into the fistula and returned to the general circulation without entering the limb's capillaries. This results in cold extremities of that limb, cramping pains, and, if severe, tissue damage. One long-term complication of an AV fistula can be the development of an aneurysm, a bulging in the wall of the vein where it is weakened by the repeated insertion of needles over time. To a large extent the risk of developing an aneurysm can be reduced by careful needling technique. Aneurysms may necessitate corrective surgery and may shorten the useful life of a fistula.To prevent damage to the fistula and aneurysm or pseudoaneurysm formation, it is recommended that the needle be inserted at different points in a rotating fashion. Another approach is to cannulate the fistula with a blunted needle, in exactly the same place. This is called a 'buttonhole' approach. Often two or three buttonhole places are available on a given fistula. This also can prolong fistula life and help prevent damage to the fistula.
=AV graft= AV (arteriovenous) grafts are much like fistulas in most respects, except that an artificial vessel is used to join the artery and vein. The graft usually is made of a synthetic material, often PTFE, but sometimes chemically treated, sterilized veins from animals are used. Grafts are inserted when the patient's native vasculature does not permit a fistula. They mature faster than fistulas, and may be ready for use several weeks after formation (some newer grafts may be used even sooner). However, AV grafts are at high risk to develop narrowing, especially in the vein just downstream from where the graft has been sewn to the vein. Narrowing often leads to clotting or thrombosis. As foreign material, they are at greater risk for becoming infected. More options for sites to place a graft are available, because the graft can be made quite long. Thus a graft can be placed in the thigh or even the neck (the 'necklace graft').Fistula First project
AV fistulas have a much better access patency and survival than do venous catheters or grafts. They also produce better patient survival and have far fewer complications compared to grafts or venous catheters. For this reason, the Centers for Medicare & Medicaid (CMS) has set up a Fistula First Initiative [ [http://fistulafirst.org/ Fistula First Initiative] ] , whose goal is to increase the use of AV fistulas in dialysis patients.
=Equipment= The hemodialysis machine pumps the patient's blood and the dialysate through the dialyzer. The newest dialysis machines on the market are highly computerized and continuously monitor an array of safety-critical parameters, including blood and dialysate flow rates; dialysis solution conductivity, temperature, and pH; and analysis of the dialysate for evidence of blood leakage or presence of air. Any reading that is out of normal range triggers an audible alarm to alert the patient-care technician who is monitoring the patient. Manufacturers of dialysis machines include companies such as Fresenius,Gambro , Baxter, B. Braun, and Bellco.Water system
An extensive
water purification system is absolutely critical for hemodialysis. Since dialysis patients are exposed to vast quantities of water, which is mixed with dialysate concentrate to form the dialysate, even trace mineral contaminants or bacterialendotoxins can filter into the patient's blood. Because the damaged kidneys cannot perform their intended function of removing impurities, ions introduced into the bloodstream via water can build up to hazardous levels, causing numerous symptoms ordeath . Aluminum, chloramine, fluoride, copper, and zinc, as well as bacterial fragments and endotoxins, have all caused problems in this regard.For this reason, water used in hemodialysis is carefully purified before use. Initially it is filtered and temperature-adjusted and its pH is corrected by adding an acid or base. Then it is softened. Next the water is run through a tank containing activated charcoal to adsorb organic contaminants. Primary purification is then done by forcing water through a membrane with very tiny pores, a so-called
reverse osmosis membrane. This lets the water pass, but holds back even very small solutes such as electrolytes. Final removal of leftover electrolytes is done by passing the water through a tank with ion-exchange resins, which remove any leftover anions or cations and replace them with hydroxyl and hydrogen molecules, respectively, leaving ultrapure water.Even this degree of water purification may be insufficient. The trend lately is to pass this final purified water (after mixing with dialysate concentrate) through a dialyzer membrane. This provides another layer of protection by removing impurities, especially those of bacterial origin, that may have accumulated in the water after its passage through the original water purification system.
Once purified water is mixed with dialysate concentrate, its conductivity increases, since water that contains charged ions conducts electricity. During dialysis, the conductivity of dialysis solution is continuously monitored to ensure that the water and dialysate concentrate are being mixed in the proper proportions. Both excessively concentrated dialysis solution and excessively dilute solution can cause severe clinical problems.
Dialyzer
The dialyzer is the piece of equipment that actually filters the blood. Almost all dialyzers in use today are of the hollow-fiber variety. A cylindrical bundle of hollow fibers, whose walls are composed of semi-permeable membrane, is anchored at each end into potting compound (a sort of glue). This assembly is then put into a clear plastic cylindrical shell with four openings. One opening or blood port at each end of the cylinder communicates with each end of the bundle of hollow fibers. This forms the "blood compartment" of the dialyzer. Two other ports are cut into the side of the cylinder. These communicate with the space around the hollow fibers, the "dialysate compartment." Blood is pumped via the blood ports through this bundle of very thin
capillary -like tubes, and the dialysate is pumped through the space surrounding the fibers. Pressure gradients are applied when necessary to move fluid from the blood to the dialysate compartment.Membrane and flux
Dialyzer membranes come with different pore sizes. Those with smaller pore size are called "low-flux" and those with larger pore sizes are called "high-flux." Some larger molecules, such as beta-2-microglobulin, are not removed at all with low-flux dialyzers; lately, the trend has been to use high-flux dialyzers. However, such dialyzers require newer dialysis machines and high-quality dialysis solution to control the rate of fluid removal properly and to prevent backflow of dialysis solution impurities into the patient through the membrane.
Dialyzer membranes used to be made primarily of cellulose (derived from cotton linter). The surface of such membranes was not very biocompatible, because exposed hydroxyl groups would activate complement in the blood passing by the membrane. Therefore, the basic, "unsubstituted" cellulose membrane was modified. One change was to cover these hydroxyl groups with acetate groups (cellulose acetate); another was to mix in some compounds that would inhibit complement activation at the membrane surface (modified cellulose). The original "unsubstituted cellulose" membranes are no longer in wide use, whereas cellulose acetate and modified cellulose dialyzers are still used. Cellulosic membranes can be made in either low-flux or high-flux configuration, depending on their pore size.
Another group of membranes is made from synthetic materials, using
polymer s such aspolyarylethersulfone ,polyamide ,polyvinylpyrrolidone , polycarbonate, and polyacrylonitrile. These synthetic membranes activate complement to a lesser degree than unsubstituted cellulose membranes. Synthetic membranes can be made in either low- or high-flux configuration, but most are high-flux.Nanotechnology is being used in some of the most recent high-flux membranes to create a uniform pore size. The goal of high-flux membranes is to pass relatively large molecules such as beta-2-microglobulin (MW 11,600 daltons), but not to pass albumin (MW ~66,400 daltons). Every membrane has pores in a range of sizes. As pore size increases, some high-flux dialyzers begin to let albumin pass out of the blood into the dialysate. This is thought to be undesirable, although one school of thought holds that removing some albumin may be beneficial in terms of removing protein-bound uremic toxins.
Membrane flux and outcome
Whether using a high-flux dialyzer improves patient outcomes is somewhat controversial, but several important studies have suggested that it has clinical benefits. The NIH-funded HEMO trial compared survival and hospitalizations in patients randomized to dialysis with either low-flux or high-flux membranes. Although the primary outcome (all-cause mortality) did not reach statistical significance in the group randomized to use high-flux membranes, several secondary outcomes were better in the high-flux group [cite journal |author=Eknoyan G, Beck GJ, Cheung AK, "et al" |title=Effect of dialysis dose and membrane flux in maintenance hemodialysis |journal=N. Engl. J. Med. |volume=347 |issue=25 |pages=2010–9 |year=2002 |pmid=12490682 |doi=10.1056/NEJMoa021583 |url=http://content.nejm.org/cgi/pmidlookup?view=short&pmid=12490682&promo=ONFLNS19] [cite journal |author=Cheung AK, Levin NW, Greene T, "et al" |title=Effects of high-flux hemodialysis on clinical outcomes: results of the HEMO study |journal=J. Am. Soc. Nephrol. |volume=14 |issue=12 |pages=3251–63 |year=2003 |pmid=14638924 |doi= |url=http://jasn.asnjournals.org/cgi/pmidlookup?view=long&pmid=14638924] . A recent Cochrane analysis concluded that benefit of membrane choice on outcomes has not yet been demonstrated [cite journal |author=Macleod AM, Campbell M, Cody JD, "et al" |title=Cellulose, modified cellulose and synthetic membranes in the haemodialysis of patients with end-stage renal disease |journal=Cochrane Database Syst Rev |volume= |issue=3 |pages=CD003234 |year=2005 |pmid=16034894 |doi=10.1002/14651858.CD003234.pub2 |url=http://www.cochrane.org/reviews/en/ab003234.html] .
Membrane flux and beta-2-microglobulin amyloidosis
High-flux dialysis membranes and/or intermittent on-line hemodiafiltration (IHDF) may also be beneificial in reducing complications of beta-2-microglobulin accumulation. Because beta-2-microglobulin is a large molecule, with a molecular weight of about 11,600 daltons, it does not pass at all through low-flux dialysis membranes. Beta-2-M is removed with high-flux dialysis, but is removed even more efficiently with IHDF. After several years (usually at least 5-7), patients on hemodialysis begin to develop complications from beta-2-M accumulation, including carpal tunnel syndrome, bone cysts, and deposits of this amyloid in joints and other tissues. Beta-2-M amyloidosis can cause very serious complications, including a spondylarthropathy, and often is associated with shoulder joint problems. Observational studies from Europe and Japan have suggested that using high-flux membranes in dialysis mode, or IHDF, reduces beta-2-M complications in comparison to regular dialysis using a low-flux membrane. [cite journal |author=van Ypersele de Strihou C, Jadoul M, Malghem J, Maldague B, Jamart J |title=Effect of dialysis membrane and patient's age on signs of dialysis-related amyloidosis. The Working Party on Dialysis Amyloidosis |journal=Kidney Int. |volume=39 |issue=5 |pages=1012–9 |year=1991 |pmid=2067196 |doi= |url= ] [ [http://www.kidney.org/professionals/kdoqi/guideline_upHD_PD_VA/hd_rec5.htm KDOQI Clinical Practice Guidelines for Hemodialysis Adequacy, 2006 Updates. CPR 5.] ] [cite journal |author=Küchle C, Fricke H, Held E, Schiffl H |title=High-flux hemodialysis postpones clinical manifestation of dialysis-related amyloidosis |journal=Am. J. Nephrol. |volume=16 |issue=6 |pages=484–8 |year=1996 |pmid=8955759 |doi= |url=] [cite journal |author=Koda Y, Nishi S, Miyazaki S, "et al" |title=Switch from conventional to high-flux membrane reduces the risk of carpal tunnel syndrome and mortality of hemodialysis patients |journal=Kidney Int. |volume=52 |issue=4 |pages=1096–101 |year=1997 |pmid=9328951 |doi= |url=] [cite journal |author=Locatelli F, Mastrangelo F, Redaelli B, "et al" |title=Effects of different membranes and dialysis technologies on patient treatment tolerance and nutritional parameters. The Italian Cooperative Dialysis Study Group |journal=Kidney Int. |volume=50 |issue=4 |pages=1293–302 |year=1996 |pmid=8887291 |doi= |url=]
Dialyzer size and efficiency
Dialyzers come in many different sizes. A larger dialyzer with a larger membrane area (A) will usually remove more solutes than a smaller dialyzer, especially at high blood flow rates. This also depends on the membrane permeability coefficient "K0" for the solute in question. So dialyzer efficiency is usually expressed as the "K0A" - the product of permeability coefficient and area. Most dialyzers have membrane surface areas of 0.8 to 2.2 square meters, and values of "K0A" ranging from about 500 to 1500 mL/min. "K0A", expressed in mL/min, can be thought of as the maximum clearance of a dialyzer at very high blood and dialysate flow rates.
Reuse of dialyzers
The dialyzer may either be discarded after each treatment or be reused. Reuse requires an extensive procedure of high-level disinfection. Reused dialyzers are not shared between patients. There was an initial controversy about whether reusing dialyzers worsened patient outcomes. The consensus today is that reuse of dialyzers, done carefully and properly, produces similar outcomes to single use of dialyzers [ [http://www.kidney.org/professionals/kdoqi/guideline_upHD_PD_VA/hd_rec5.htm KDOQI Clinical Practice Guidelines for Hemodialysis Adequacy, 2006 Updates] . CPR 5.] .
ee also
*
Dialysis
*Dialysis disequilibrium syndrome
*Home hemodialysis
*Peritoneal dialysis
*Hemofiltration
*Extracorporeal therapies
*Renal replacement therapy
*Step-by-step description of hemodialysis References
External links
* [http://kidney.niddk.nih.gov/kudiseases/pubs/yourkidneys/ Your Kidneys and How They Work] - (American) National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK),
NIH .
* [http://kidney.niddk.nih.gov/kudiseases/pubs/kidneyfailure/index.htm Treatment Methods for Kidney Failure] - (American) National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK),NIH .
* [http://kidney.niddk.nih.gov/kudiseases/pubs/hemodialysis/ Treatment Methods for Kidney Failure: Hemodialysis] - (American) National Kidney and Urologic Diseases Information Clearinghouse,NIH .