- Cardiopulmonary bypass
Cardiopulmonary bypass (CPB) is a technique that temporarily takes over the function of the
heart andlung s during surgery, maintaining the circulation of blood and the oxygen content of the body. The CPB pump itself is often referred to as a "Heart-Lung Machine" or "the Pump". Cardiopulmonary bypass pumps are operated byallied health professional s known as perfusionists in association with surgeons who connect the pump to the patient's body. CPB is a form ofextracorporeal circulation.Uses of cardiopulmonary bypass
Cardiopulmonary bypass is commonly used in
heart surgery because of the difficulty of operating on the beating heart. Operations requiring the opening of the chambers of the heart require the use of CPB to support the circulation during that period.CPB can be used for the induction of total body
hypothermia , a state in which the body can be maintained for an hour or more withoutperfusion (blood flow). If blood flow is stopped at normalbody temperature , permanentbrain damage normally occurs in three to four minutes — death may follow shortly afterward.Extracorporeal membrane oxygenation (
ECMO ) is a simplified form of CPB sometimes used aslife-support for newborns with seriousbirth defects , or tooxygenate and maintain recipients fororgan transplantation until new organs can be found.CPB mechanically circulates and oxygenates blood for the body while bypassing the heart and lungs. It uses a heart-lung machine to maintain perfusion to other body organs and tissues while the surgeon works in a bloodless surgical field. The surgeon places a
cannula in right atrium, vena cava, or femoral vein to withdraw blood from the body. The cannula is connected to tubing filled with isotonic crystalloid solution. Venous blood that is removed from the body by the cannula is filtered, oxygenated, cooled or warmed, and then returned to the body. The cannula used to return oxygenated blood is usually inserted in the ascending aorta, but it may be inserted in the femoral artery. Pt receives heparin to prevent thrombus formation, and protamine sulfate given after to reverse effects of heparin. During procedure, hypothermia maintained; body temperature usually 28ºC to 32ºC (82.4-89.6ºF). Blood is cooled during CPB and returned to the body. The cooled blood slows the body’s basal metabolic rate, decreasing its demand for oxygen. Cooled blood usually has a higher viscosity, but the crystalloid solution used to prime the bypass tubing dilutes the blood.urgical procedures in which cardiopulmonary bypass is used
*
Coronary artery bypass surgery
*Cardiac valve repair and/or replacement (aortic valve ,mitral valve ,tricuspid valve ,pulmonic valve )
*Repair of large septal defects (atrial septal defect ,ventricular septal defect ,atrioventricular septal defect )
*Repair and/or palliation ofcongenital heart defect s (Tetralogy of Fallot ,transposition of the great vessels )
*Transplantation (heart transplantation ,lung transplantation ,heart-lung transplantation )
*Repair of some largeaneurysms (aortic aneurysm s,cerebral aneurysm s)
*Pulmonary thromboendarterectomy
*Pulmonary thrombectomy History
Dr. Clarence Dennis led the team that conducted the first known operation involving open cardiotomy with temporary mechanical takeover of both heart and lung functions on April 5, 1951 at the University of Minnesota Hospital. The patient did not survive due to an unexpected complex congenital heart defect. This followed four years of laboratory experimentation with dogs. [Dennis C, Spreng DS, Jr., Nelson GE, et al. [http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=1802968 Development of a pump-oxygenator to replace the heart and lungs; an apparatus applicable to human patients, and application to one case.] Ann Surg 1951; 134:709-721]
The first successful open heart procedure on a human utilizing the heart lung machine was performed by John Gibbon on May 6, 1953 in
Philadelphia . He repaired anatrial septal defect in an 18-year-old woman. [Cohn, Lawrence. Fifty Years of Open-Heart Surgery. Circulation. 2003;107:2168]Components of cardiopulmonary bypass
Cardiopulmonary bypass consists of two main functional units, the
pump and theoxygenator which remove oxygen-deprived blood from a patient's body and replace it with oxygen-rich blood through a series of hoses.Tubing
The components of the CPB circuit are interconnected by a series of tubes made of
silicone rubber , orPVC . The tubing in the CPB circuit is similar to transparentgarden hose .Pumps
Roller pump
The pump console usually comprises several rotating motor-driven pumps that peristaltically "massage" tubing . This action gently propels the blood through the tubing. This is commonly referred to as a roller pump, or
peristaltic pump .Centrifugal pump
Many CPB circuits now employ a
centrifugal pump for the maintenance and control of blood flow during CPB. By altering the speed of revolution (RPM) of the pump head, blood flow is produced by centrifugal force. This type of pumping action is considered to be superior to the action of the roller pump by many because it is thought to produce less blood damage (Hemolysis , etc.).Oxygenator
The
oxygenator is designed to transferoxygen to infusedblood and removecarbon dioxide from thevenous blood . Cardiac surgery was made possible by CPB usingbubble oxygenator s, butmembrane oxygenator s have supplanted bubble oxygenators since the 1980s.The oxygenator was first conceptualised in the 17th century by
Robert Hooke and developed into practical extracorporeal oxygenators by French and German experimental physiologists in the 19th century. Bubble oxygenators have no intervening barrier between blood and oxygen, these are called 'direct contact' oxygenators. Membrane oxygenators introduce a gas-permeable membrane between blood and oxygen that decreases the blood trauma of direct-contact oxygenators. Much work since the 1960s focused on overcoming the gas exchange handicap of the membrane barrier, leading to the development of high-performance microporous hollow-fibre oxygenators that eventually replaced direct-contact oxygenators in cardiac theatres.cite journal |author=Lim M |title=The history of extracorporeal oxygenators |journal=Anaesthesia |volume=61 |issue=10 |pages=984–95 |year=2006 |pmid=16978315 |doi=10.1111/j.1365-2044.2006.04781.x]Another type of oxygenator gaining favour recently is the
heparin-coated blood oxygenator which is believed to produce less systemic inflammation and decrease the propensity for blood to clot in the CPB circuit.Cannulae
Multiple cannulae are sewn into the patient's body in a variety of locations, depending on the type of surgery. A venous cannula removes oxygen deprived blood from a patient's body. An arterial cannula is sewn into a patient's body and is used to infuse oxygen-rich blood. A
cardioplegia cannula is sewn into the heart to deliver a cardioplegia solution to cause the heart to stop beating.Cardioplegia
A CPB circuit consists of a systemic circuit for oxygenating blood and reinfusing blood into a patient's body (bypassing the heart); and a separate circuit for infusing a solution into the heart itself to produce
cardioplegia (i.e. to stop the heart from beating), and to providemyocardial protection (i.e. to prevent death of heart tissue).Operation
A CPB circuit must be primed with fluid and all air expunged before connection to the patient. The circuit is primed with a crystalloid solution and sometimes blood products are also added. The patient must be fully anticoagulated with an
anticoagulant such asheparin to prevent massive clotting of blood in the circuit.Complications
CPB is not benign and there are a number of associated problems:
*
Postperfusion syndrome (also known as Pumphead)
*Hemolysis
*Capillary Leak Syndrome
*Clotting of blood in the circuit - can block the circuit (particularly the oxygenator) or send a clot into the patient.
*Air embolism
*Leakage - a patient can rapidlyexsanguinate (lose blood perfusion of tissues) if a line becomes disconnected.References
External links
* [http://cardiacsurgery.ctsnetbooks.org/cgi/content/full/2/2003/317 Extracorporeal Circulation: Perfusion Systems. In: Cohn LH, Edmunds LH Jr, eds. Cardiac Surgery in the Adult]
* [http://mmcts.ctsnetjournals.org/cgi/collection/cardiopulmonary_bypass Multimedia Manual of Cardiothoracic Surgery. Cardiopulmonary bypass collection.]
* [http://perfline.com/index.htm Perfusion Line]
* [http://perfline.com/textbook/textbook.html The Virtual Textbook Of Extracorporeal Technology]
* [http://video.google.com/videoplay?docid=4187970374764805675 Video of early USSR heart-lung machine experiments]
Wikimedia Foundation. 2010.