- Absolute molar mass
Absolute Molar Mass is a process to determine the characteristics of molecules.
__TOC__
History
The first
absolute measurements (i.e. made without reference to standards) were based onfundamental physical characteristic s and their relation to the molar mass. The most useful of these were Membrane Osmometry andSedimentation .Another absolute instrumental approach was also possible with the development of
light scattering byEinstein ,Raman ,Debye ,Zim , and others. The problem with measurements made using membrane osmometry and sedimentation was that they only characterized the bulk properties of thepolymer sample. Moreover, the measurements were excessively time consuming and prone tooperator error . In order to gain information about apolydisperse mixture of molar masses, a method for separating the different sizes was developed. This was achieved by the advent of High Performance Size Exclusion Chromatography (HPSEC). HPSEC is based on the fact that the pores in the packing material of chromatography columns could be made small enough for molecules to become temporarily lodged in their interstitial spaces. As the sample makes its way through a column the smaller molecules spend more time traveling in these void spaces than the larger ones, which have fewer places to "wander". The result is that a sample is separated according to itshydrodynamic volume . As a consequence, the big ones come out first, and then the small ones elute. By choosing a suitable column packing material it is possible to define the resolution of the system. Columns can also be combined in series to increase resolution or the range of sizes studied.The next step is to convert the time at which the samples eluted into a measurement of molar mass. This is possible because if the molar mass of a standard were known the time at which this standard eluted should be equal to a specific molar mass. Using multiple standards, a
calibration curve of time versus molar mass was developed. This is significant for polymer analysis because a single polymer could be shown to be comprised of many different components, and the complexity and distribution of which would also affect the physical properties. However this technique has shortcomings. For example, unknown samples were always measured in relation to known standards, and these standards may or may not have had any similarities to the sample of interest. The measurements made by HPSEC were then mathematically converted into data similar to that found by the existing techniques.The problem was that the system was calibrated according to the Vh characteristics of polymer standards that are not directly related to the molar mass. If the relationship between the molar mass and Vh of the standard is not the same as that of the unknown sample, then the calibration is invalid. Thus, to be accurate, the calibration must use the same polymer, of the same conformation, in the same eluent and have the same interaction with the solvent as the hydration layer changes Vh. Moreover, there must be no interaction with the column other than that produced by size. As the demands on polymer properties increased, the necessity of getting absolute information on the molar mass and size also increased. This was especially important in pharmaceutical applications where slight changes in
molar mass (e.g.aggregation ) or shape may result in differentbiological activity . These changes can actually have a harmful effect instead of a beneficial one.Traditional
light scattering instruments worked by taking readings from multiple angles, each being measured in series. A low angle light scattering system8 was developed in the early 1970s that allowed a single measurement to be used to calculate the molar mass. Although measurements at low angles are problematic for fundamental physical reasons (molecules tend to scatter more light in lower angle directions than in higher angles, and low angle scattering events caused by dust and contamination of the mobile phase easily overwhelm the scattering from the molecules of interest), low-angle laser light scattering (LALLS) became popular in the 1970s and mid-1980s.Multi-angle light scattering was invented in the mid-1980s and instruments like instruments9 that were able to make measurements at the different angles simultaneously but it was not until the later 1980s (10-12) that the connection of
multi-angle laser light scattering (MALS) detectors to SEC systems was a practical proposition enabling both molar mass and size to be determined from each slice of the polymer fraction.Multi Angle (Laser) Light Scattering (MALS)
MALS can be applied to
synthetic polymer s,protein s,pharmaceuticals and particles such asliposome s,micelles , and encapsulated proteins. Measurements can be made in one of two modes which areun-fractionated (batch mode) or incontinuous flow mode (with HPSEC, HPLC or any otherflow fractionation method ). Batch mode experiments can be performed either by injecting a sample into a flow cell with a syringe or with the use of discrete vials. These measurements are most often used to measure timed events likeantibody-antigen reactions orprotein assembly . Batch mode measurements can also be used to determine the second virial coefficient (A2), a value that gives a measure of the likelihood of crystallization or aggregation in a given solvent. Continuous flow experiments can be used to study material eluting from virtually any source. More conventionally, the detectors are coupled to a variety of different chromatographic separation systems. The ability to determine the mass and size of the materials eluting then combines the advantage of the separation system with an absolute measurement of the mass and size of the species eluting.MALS measurements work by calculating the amount of light scattered at each angle detected. This process overcomes the problems associated with low angle detectors (typically there is around ten times the noise at an angle of 11º or below compared to 90º) and allows a reliable and accurate measure of the light scattered. The higher the number of detectors, the better the accuracy of the experiment. The amount of light scattered is then related to themolar mass.
The addition of a MALS detector coupled downstream to a chromatographic system allows the utility of SEC or similar separation combined with the advantage of an absolute detection method. The light scattering data is purely dependent on the light scattering signal times the concentration; the elution time is irrelevant and the separation can be changed for different samples without recalibration. In addition, a non-size separation method such as HPLC or IC can also be used. As the light scattering detector is mass dependent, it becomes more sensitive as the molar mass increases. Thus it is an excellent tool for detecting aggregation. The higher the aggregation number, the more sensitive the detector becomes. As previously noted, the MALS detector can also provide information about the size of the molecule. This information is the Root Mean Square radius of the molecule (RMS or Rg). This is different from the Rh mentioned above, in that it is not affected by the hydration layer and is purely the root mean square of all the radii making up the molecule multiplied by the mass at that radius.
This would seem like a strange parameter to measure but in fact this is very useful as it is sensitive to changes in shape of the molecule. If you consider adding a side group to a large molecule (branching), this would hardly affect the Rh but would have a significant impact on the Rg. In addition, if you plot the Log Rg vs. Log M then the conformation of the molecule can be derived.
Bibliography
*A. Einstein, Ann. Phys. 33 (1910), 1275
*C.V. Raman, Indian J. Phys. 2 (1927), 1
*P.Debye, J. Appl. Phys. 15 (1944), 338
*B.H. Zimm, J. Chem. Phys 13 (1945), 141
*B.H. Zimm, J. Chem. Phys 16 (1948), 1093
*B.H. Zimm, R.S. Stein and P. Dotty, Pol. Bull. 1,(1945), 90
*M. Fixman, J. Chem. Phys. 23 (1955), 2074
*A.C. Ouano and W. Kaye J. Poly. Sci. A1(12) (1974), 1151
*Flow Through MALS detector, DLS 800, Science Spectrum Inc.
*P.J. Wyatt, C. Jackson and G.K. Wyatt Am. Lab 20(6) (1988), 86
*P.J. Wyatt, D. L. Hicks, C. Jackson and G.K. Wyatt Am. Lab. 20(6) (1988), 106
*C. Jackson, L.M. Nilsson and P.J. Wyatt J. Appl. Poly. Sci. 43 (1989), 99
Wikimedia Foundation. 2010.