- Fluorescein angiography
Fluorescein angiography , or fluorescent angiography, is a technique for examining the circulation of theretina using thedye tracing method. It involves injection of sodiumfluorescein into the systemic circulation, and then anangiogram is obtained by photographing thefluorescence emitted after illumination of the retina with blue light at awavelength of 490nanometer s. The fluorescein dye also reappears 12-24 hours in the patient urine, causing a yellow-green appearance.Fluorescein angiography is one of several
health care applications of this dye, all of which have a risk of severeadverse effect s. See fluorescein safety in health care applications.Equipment
*Exciter filter: allows only blue light to illuminate the retina. It reduces the amount of non-fluorescent light that can reach the film, and allows for maximum excitation of the fluorescein.
*Barrier filter : allows only yellow-green light (from the fluorescence) to reach the camera. It absorbs the blue light used to illuminate the retina.
*Camera with black and white film. Also digital cameras have come into use since the late 1990s.Technique
*Baseline color and red-free filtered images are taken prior to injection. This allows for the recognition of autofluorescence of the retinal tissues.
*A 6 second bolus Injection of approximately 5cc of sodium fluorescein into a vein in the arm or hand.
*A series of black-and-white or digital photographs are taken of the retina after the fluorescein reaches the retinal circulation (approximately 10 seconds after injection). Photos are taken approximately once every second for about 20 seconds, then less often. A delayed image is obtained at 5 and 10 minutes. Some doctors like to see a 15 minute image as well.
*A filter is placed in the camera so only the fluorescent, yellow-green light (530 nm) is recorded. The camera may however pick up signals frompseudofluorescence orautofluorescence . In pseudofluorescence, non-fluorescent light is imaged. This occurs when blue light reflected from the retina passes through the filter. This is generally a problem with older filters, and annual replacement of these filters is recommended. In autofluorescence, fluorescence from the eye occurs without injection of the dye. This may be seen with [http://www.revoptom.com/handbook/SECT50a.HTM optic nerve head drusen] , [http://www.revoptom.com/handbook/sect6l4b.htm astrocytic hamartoma] , or calcific scarring.
*Black-and-white photos give better contrast than color photos, which aren't necessary since only one color is being transmitted though the filter.Normal circulatory filling
"times are approximate"
*0 seconds — injection of fluorescein
*9.5 sec — posterior ciliary arteries
*10 sec —choroid al flush (or "pre-arterial phase")
*10-12 sec — retinal arterial stage
*13 sec —capillary transition stage
*14-15 sec — early venous stage (or "lamellar stage", "arterial-venous stage")
*16-17 sec — venous stage
*18-20 sec — late venous stage
*5 minutes — late stainingFluorescein enters the ocular circulation from the
internal carotid artery via theophthalmic artery . The ophthalmic artery supplies the choroid via the short posterior ciliary arteries and the retina via the central retinal artery, however, the route to the choroid is typically less circuitous than the route to the retina. This accounts for the short delay between the "choroidal flush" and retinal filling.Pathologic findings
Pathologic changes are recognized by the detection of either hyperfluorescence or hypofluroescence.
Causes of hyperfluorescence::leaking (i.e. capillary leakage,
aneurysm ,neovascularization ):pooling (serous retinal detachment):staining:transmission defects:abnormal vessels Causes of hypofluorescence::optical barrier (i.e. blood):filling defect (capillary blockage)Among the common groups of ophthalmologic disease, fluorescein angiography can detect
diabetic retinopathy , vein occlusions, retinal artery occlusions, edema of the optic disc, and tumors.fluorescence angiography is not the most preferred technique as it has many disadvantages like discomfort to the patient.Other types of fluorescent angiography
*Indocyanine Green Angiography
References
* [http://www.nlm.nih.gov/medlineplus/ency/article/003846.htm Fluorescein angiography in Medical Encyclopedia]
ee also
*
Fundus camera
Wikimedia Foundation. 2010.