- Intra-aortic balloon pump
The Intra-aortic balloon pump (IABP) is a mechanical device that is used to decrease myocardial oxygen demand while at the same time increasing
cardiac output . By increasing cardiac output it also increases coronary blood flow and therefore myocardial oxygen delivery. It consists of a cylindrical balloon that sits in theaorta and counterpulsates. That is, it actively deflates insystole increasing forward blood flow by reducingafterload thus, and actively inflates indiastole increasing blood flow to the coronary arteries. These actions have the combined result of decreasing myocardial oxygen demand and increasing myocardial oxygen supply. [http://www.lww.com/product/?0-7817-3548-3 Intensive Care Medicine] by Irwin and Rippe] [http://www.rjmatthewsmd.com/Definitions/IABP_Counterpulsation.htm Intra Aortic Balloon Pump (IABP) Counterpulsation] [http://www.uam.es/departamentos/medicina/anesnet/journals/ijeicm/vol3n2/iabp.htm mirror with better quality] by P. J Overwalder, M.D., Department of Surgery, Division of Cardiac Surgery, University Hospital Graz, The Internet Journal of Thoracic and Cardiovascular Surgery. 1999. Volume 2 Number 2.] [http://www.aic.cuhk.edu.hk/web8/IABP.htm Intra-aortic balloon pumping] Department of Anaesthesia and Intensive Care of The Chinese University ofHong Kong ] The balloon is inflated during diastole by a computer controlled mechanism, usually linked to either an ECG or a pressure transducer at the distal tip of the catheter; some IABPs, such as the Datascope System 98XT, allow for asynchronous counterpulsation at a set rate, though this setting is rarely used. The computer controls the flow of helium from a cylinder into and out of the balloon. Helium is used because its low viscosity allows it to travel quickly through the long connecting tubes, and has a lower risk of causing a harmful embolism should the balloon rupture while in use.History
The IABP device was pioneered at the
Grace-Sinai Hospital in Detroit during the early 1960s by Dr. Adrian Kantrowitz.Use of counterpulsation
Intraaortic balloon counterpulsation is used in situations when the heart's own cardiac output is insufficient to meet the oxygenation demands of the body. These situations could include cardiogenic shock, severe septic shock, post cardiac surgery and numerous other situations.
Indications
The following situations may benefit from this device.
*Cardiogenic shock when used alone as treatment formyocardial infarction 9-22% survive the first year.
*Reversible intracardial mechanical defects complicating infarction, i.e. acute mitral regurgitation and septal perforation.
*Unstableangina pectoris benefits from counterpulsation.
*Postcardiothoracic surgery most common and useful is the use of counterpulsation in weaning patients from cardiopulmonary bypass after continued perioperative injury to myocardial tissue.
*Preoperative use has been suggested for high-risk patients such as those with unstable angina with stenosis greater than 70% of main coronary artery, in ventricaular dysfunction with anejection fraction less than 35%.
*Bridge toheart transplant for those patients with left ventricular failure.
*Percutaneous coronary angioplastyContraindications
Absolute contraindication
The following conditions will always exclude patients for treatment:
*Aortic valve insufficiency
*Aortic dissection
*Severe aortoiliac occlusive diseaseRelative contraindication
The following conditions could, under very pressing circumstances, be allowed to be included for therapy:
*Prosthetic vascular grafts in theaorta
*Aorticaneurysm
*Aortofemoral graftsComplications
Since the device is placed in the
femoral artery and aorta it could provokeischemia , andcompartment syndrome . At highest risk is the leg which is supplied by the femoral artery may become ischemic, but also placing the balloon too distal from the arcus aortae may induce occlusion of therenal artery and subsequentrenal failure . Other possible complications are cerebral embolism during insertion, infection,dissection of the aorta oriliac artery , perforation of the artery and hemorrhage in themediastinum . Mechanical failure of the balloon itself is also a risk which entails vascular surgery to remove under that circumstance. After balloon removal there is also a risk of 'embolic shower' from micro clots that have formed on the surface of the balloon, and can lead to peripheral thrombosis, myocardial ischemia, hemodynamic decompensation, and late pseudoaneurysm.Trivia
The IABP's developers thought that private insurance companies would not stand for such a radical new therapy, and approached the military, asking them to keep an eye out for IABP candidates in military hospitals. One day, the call came. A retired general was hospitalized in Walter Reed Army Medical Center. He had a history of multiple infarcts and was now end-stage heart failure. The IABP scientists were all ready to go, until they found out the general was former President
Dwight D. Eisenhower . They declined the opportunity because they worried that if the IABP failed in him, it would permanently ruin the future prospects of the device. Fact|date=August 2007ee also
*
Cardiogenic shock
*Ventricular assist device References
External links
* [http://texasheart.org/Research/Devices/iabp.cfm Intra-aortic Balloon Pump (IABP)] Texas Heart Institute
* [http://people.brunel.ac.uk/~emstawk/IABP.htm Intra Aortic Balloon Pump (IABP)] two videos demonstrating inflating and deflating the balloon
* [http://www.italheartj.org/pdf_files/20050080.pdf Images in cardiovascular medicine The missing intra-aortic balloon pump catheter] by Pasquale Totaro, Nello Degno, John Smith, Vincenzo Argano, Cardiac Surgery Department, Regional Cardiac Center, Morriston Hospital, Swansea, UK, Ital Heart J 2005; 6 (4): 361-362)