- Supercritical angle fluorescence microscopy
Supercritical Angle Fluorescence Microscopy (SAF) is a technique to detect and characterize fluorescent species (proteins, biomolecules, pharmaceuticals, etc.) and their behaviour very close or even adsorbed or linked at surfaces. The method is able to observe molecules in a distance of less than 100 to 0 nanometer from the surface even in presence of high concentrations of fluorescent species around. Using an aspheric lens for excitation of a sample with laser light,
fluorescence emitted by the specimen is collected above thecritical angle oftotal internal reflection selectively and directed by a parabolic optics onto a detector. The method was invented in 1998 in the laboratories ofStefan Seeger atUniversity of Regensburg /Germany and later atUniversity of Zurich /Switzerland.SAF Microscopy Principle
The principle how SAF Microscopy works is as follows: A fluorescent specimen does not emit fluorescence isotropically when it comes close to a surface, but approximately 70% of the fluorescence emitted is directed into the solid phase. Here, the main part enters the solid body above the critical angle [J. Enderlein, T. Ruckstuhl, S. Seeger: Highly Efficient Optical Detection of Surface-Generated Fluorescence. Appl. Opt. 38 (4) 724-32 (1999)] . When the emitter is located just 200 nm above the surface, fluorescent light entering the solid body above the critical angle is decreased dramatically. Hence, SAF Microscopy is ideal suited to discriminate between molecules and particles at or close to surfaces and all other specimen present in the bulk [T. Ruckstuhl, M. Rankl, S. Seeger: Highly sensitive biosensing using a Supercritical Angle Fluorescence (SAF) instrument, Biosensors&Bioelectronics 18 (9) 1193-1199 (2003)] , [T. Ruckstuhl, S. Seeger: Attoliter detection volumes by confocal total-internal-reflection fluorescence microscopy, Optic Letters 29, 569-571 (2004)] .
Typical SAF-setup
The typical SAF setup consists of a laser line (typically 450-633 nm), which is reflected into the aspheric lens by a dichroic mirror. The lens focuses the laser beam in the sample, causing the particles to fluoresce. The fluorescent light then passes through a parabolic lens before reaching a detector, typically a
photomultiplier tube oravalanche photodiode detector.Selected publications