- Photoplethysmograph
A photoplethysmograph (PPG) is an optically obtained
plethysmograph , a volumetric measurement of an organ. A PPG is often obtained by using apulse oximeter which illuminates theskin and measures changes in light absorption (Shelley and Shelley, 2001). A conventional pulse oximeter monitors the perfusion of blood to the dermis and subcutaneous tissue of the skin.With eachcardiac cycle the heart pumps blood to the periphery. Even though this pressure pulse is somewhat damped by the time it reaches the skin, it is enough to distend the arteries and arterioles in the subcutaneous tissue. If the pulse oximeter is attached without compressing the skin, a pressure pulse can also be seen from the venous plexus, as a small secondary peak.The change in volume caused by the pressure pulse is detected by illuminating the skin with the light from a
Light Emitting Diode (LED) and then measuring the amount of light either transmitted or reflected to a photodiode. Each cardiac cycle appears as a peak, as seen in the figure. Because blood flow to the skin can be modulated by multiple other physiological systems, the PPG can also be used to monitor breathing,hypovolemia , etc. Additionally, the shape of the PPG waveform differs from subject to subject, and varies with the location and manner in which the pulse oximeter is attached.Sites for measuring PPG
While
pulse oximeter s are a commonly used medical device the PPG derived from them is rarely displayed, and is nominally only processed to determine heart rate. PPGs can be obtained from transmissive absorption (as at the finger tip) or reflective (as on the forehead).In outpatient setting pulse oximeters are commonly worn on the finger and ear. However, in cases of shock,
hypothermia , etc. blood flow to the periphery can be reduced, resulting in a PPG without a discernible cardiac pulse. In this case, a PPG can be obtained from a pulse oximeter on the head, with the most common sites being the ear,nasal septum , and forehead.PPGs can also be obtained from the vagina and
esophagus .Uses
Monitoring Heart Rate and Cardiac Cycle
Because the skin is so richly perfused, it is relatively easy to detect the pulsatile component of the cardiac cycle. The DC component of the signal is attributable to the bulk absorption of the skin tissue, while the AC component is directly attributable to variation in blood volume in the skin caused by the pressure pulse of the cardiac cycle.
The height of AC component of the photoplethysmogram is proportional to the pulse pressure, the difference between the systolic and diastolic pressure in the arteries. As seen in the figure showing Premature Ventricular Contractions (PVCs)the PPG pulse for the cardiac cycle with the PVC results in lower amplitude
blood pressure and a PPG. Ventricular Tachycardia and Ventricular Fibrillation can also be detected.Monitoring Respiration
Respiration effects the cardiac cycle by varying the intrapleural pressure, the pressure between the thoracic wall and the lungs. Since the heart resides in the thoracic cavity between the lungs, the partial pressure of inhaling and exhaling greatly influence the pressure on the vena cava and the filling of the right atrium. This effect is often referred to as normal sinus arrhythmia.
During inspiration, intrapleural pressure decreases by up to 4 mm Hg which distends the right atrium, allowing for faster filling from the vena cava, increasing ventricular preload, and increasing the stroke volume. Conversely during expiration, the heart is compressed, decreasing cardiac efficiency and reducing stroke volume. However, the overall net effect of respiration is to act as pump for the cardiovascular system. When the frequency and depth of respiration increases, the venous return increase leading to increased cardiac output. (Shelley, et al., 2006)
Monitoring Depth of Anesthesia
Anesthesiologist must often judge subjectively whether a patient is sufficiently anesthetized for surgery. As seen in the figure if a patient is not sufficiently anesthetized the sympathetic nervous system response to an incision can generate an immediate response in the amplitude of the PPG.
Monitoring Hypo- and Hyper-volemia
Shamir, Eidelman, et al. studied the interaction between inspiration and removal of 10% of a patient’s blood volume for blood banking before surgery (Shamir, Eidelman et al. 1999). They found that blood loss could be detected both from the photoplethysmogram from a pulse oximeter and an arterial catheter. Patients showed a decrease in the cardiac pulse amplitude caused by reduced cardiac preload during exhalation when the heart is being compressed.
References
# M. Shamir, L. A. Eidelman, Y. Floman, L. Kaplan, and R. Pi-zov, "Pulse Oximetry Plethysmographic Waveform During Changes in Blood Volume", Br. J. Anaesth., vol. 82, pp. 178-181, 1999.
# K. Shelley and S. Shelley, "Pulse Oximeter Waveform: Photoelectric Plethysmography",in Clinical Monitoring, Carol Lake, R. Hines, and C. Blitt, Eds.: W.B. Saunders Company, 2001, pp. 420-428.
# K. H. Shelley, D. H. Jablonka, A. A. Awad, R. G. Stout, H. Rezkanna, and D. G. Silverman, "What Is the Best Site for Measuring the Effect of Ventilation on the Pulse Oximeter Waveform?" Anesth Analg, vol. 103, pp. 372-377, 2006.External links
[http://www.swarthmore.edu/NatSci/echeeve1/Ref/E72Projects/E72Cardio/E72Cardio.html A student project: building a device for collecting PPGs]