- Beam's eye view
Beam's Eye View (or BEV) is an imaging technique used in
radiation therapy for the quality assurance and planning of External Beam Radiation Therapy treatments. These are primarily used to ensure that the relative orientation of the patient and the treatment machine are correct. The BEV image will typically contain the images of the patient's anatomy and the beam modifiers (such as jaws or Multi-Leaf Collimators (or MLCs)).Generation of Beam's Eye Views
* Physical Construction:
** BEV's can be generated by exposing a high energy film (similar tophotographic film ) or an Electronic Portal Imaging Device (EPID) with the treatment beam itself after it passes through the patient and any beam modifiers (such as blocks). Although this type of image is an excellent indication of the basic quality of the treatment plan, the quality of film images can be poor.
** A BEV can be created using a radiation therapy simulator which mimics the treatment geometry (couch angle, gantry angle, etc.) using anX-ray source instead of the higher energy treatment source. The jaws and blocks can be imaged on the same film as the patient's landmarks.
* Artificial Reconstruction: The BEV can be created using aDigitally Reconstructed Radiograph (or DRR) that is created from a computed tomography (or CT) data set. This image would contain the same treatment plan information, but the patient image is reconstructed from the CT image data using a physics model.References
* Faiz Kahn and Roger Potish (Eds.) (1998)."Treatment Planning in Radiation Oncology". Williams & Wilkins. ISBN 0-683-04607-1.
* Jacob Van Dyk (Ed.) (1999). "The Modern Technology of Radiation Oncology". Medical Physics Publishing. ISBN 0-944838-38-3.
Wikimedia Foundation. 2010.