- Enterotube II
The Enterotube II is an example of a rapid, multi test system used in identification of unknown
oxidase - negative, gram- negative, rod shaped bacteria of the family "Enterobacteriaceae ". cite journal
author=L J Hayek and G W Willis
year=1984
url=http://www.pubmedcentral.nih.gov/articlerender.fcgi?tool=pmcentrez&artid=498712
title=Identification of the Enterobacteriaceae: a comparison of the Enterotube II with the API 20E |journal=J Clin Pathol.
month=March
volume=37(3)
pages=344–347
pmid=777035
doi=10.1136/jcp.37.3.344 ] It consists of a tube with a flat side and contains 12 compartments for differentbiochemical tests. Although the manufacturer is continuously improving the accuracy of this system, it is worth noting that this system may sometimes yield false results.Methods and Procedure
The Enterotube II is used by first removing the caps from both ends to expose the
inoculation wire. The wire is sterile and need not be flamed. The Enterotube II is inoculated by touching the wire to a well isolated colony from aPetri plate . The wire is pulled and rotated from the other end to inoculate all the compartments and pushed all the way back in to reinoculate the compartments. The wire is then pulled until it reaches the indole compartment and is broken off using a pair of pliers. The perforation on the aerobic compartments must be punctured using a flamed inoculation needle or a similar device. The tube is finally recapped and incubated at 37 degrees Celsius for 24 hours.Reading the Enterotube II Results
After 24 hours of incubation, any color change is recorded in the ID card provided by the manufacturer or from a separate vendor. The appropriate reagents are added to the indole and
Voges- Proskauer compartments and their results are recorded in the same ID card. The numerical values of the "positive" tests are added in their appropriate sections to yield a 5-digit ID for the organism being tested. This 5-digit number is looked up in a reference book or computer software to determine the identity of the bacteria.Accuracy and Limitations
Although the manufacturer has improved on the accuracy of the Enterotube II since it was first produced in the early 1970’s, it must be noted that the identification results is not always correct and may require retesting. In one study, the first testing of the Enterotube II correctly identified only 90.6% of the unknown bacterial cultures. Repeat testing using the Enterotube II identified 97.9% of the bacterial cultures (a total of 235 organisms were used).
References
ee also
API 20E system for identifying bacteria of the family "Enterobacteriaceae "External links
* [http://www.bd.com/ds/productCenter/IS-EnterotubeOxiFerm.asp Becton, Dickinson and Company website]
* [http://www.bdregdocs.com/msds/view/msds-usa-211832.pdf MSDS ]----Enterotube II is a trademark of
Becton, Dickinson and Company
Wikimedia Foundation. 2010.