Transcutaneous pacing

Transcutaneous pacing

Transcutaneous pacing (also called external pacing) is a temporary means of pacing a patient's heart during a medical emergency. It is accomplished by delivering pulses of electric current through the patient's chest, which stimulates the heart to contract.

The most common indication for transcutaneous pacing is an abnormally slow heart rate. By convention, a heart rate of less than 60 beats per minute in the adult patient is called bradycardia. Not all bradycardias require medical treatment. Many athletes have a slow resting heart rate. In addition, the heart rate is known to slow down with age. It is only when bradycardia causes signs and symptoms of shock that it requires emergency treatment with transcutaneous pacing.

Some common causes of hemodynamically significant bradycardia include acute myocardial infarction, sinus node dysfunction and complete heart block. Transcutaneous pacing is no longer indicated for the treatment of asystole (cardiac arrest associated with a "flat line" on the ECG), with the possible exception of witnessed asystole (as in the case of bifascicular block that progresses to complete heart block without an escape rhythm).

During transcutaneous pacing, pads are placed on the patient's chest, either in the anterior/lateral position or the anterior/posterior position. The anterior/posterior position is preferred as it minimizes transthoracic impedance by "sandwiching" the left ventricle between the two pads. The pads are then attached to a monitor/defibrillator, a heart rate is selected, and current (measured in milliamps) is increased until electrical capture (characterized by a wide QRS complex with tall, broad T wave on the ECG) is obtained, with a corresponding pulse. Pacing artifact on the ECG and severe muscle twitching may make this determination difficult. It is therefore advisable to use another instrument (e.g. SpO2 monitor or bedside doppler) to confirm mechanical capture.

Transcutaneous pacing may be uncomfortable for the patient. Sedation should therefore be considered. Prolonged transcutaneous pacing may cause burns on the skin. It is meant to stabilize the patient until a more permanent means of pacing is achieved.

Other forms of cardiac pacing are transvenous pacing, epicardial pacing, and permanent pacing with an implantable pacemaker.

References

# Urden, L., Stacy, K., and Lough, M. "Thelan's Critical Care Nursing: Diagnosis and Management". Fourth Edition, Mosby, 1998. ISBN 0-323-01461-5
# "Handbook of Emergency Cardiovacular Care for Healthcare Providers". Editors Hazinski, M., Cummins, R., and Field, J. 2004. ISBN 0-87493-448-6


Wikimedia Foundation. 2010.

Игры ⚽ Нужно решить контрольную?

Look at other dictionaries:

  • transcutaneous pacing — a temporary method for cardiac pacing, in which large surface, high impedance electrodes are applied to the anterior and posterior chest walls to deliver high current stimuli of long duration for pacing of the ventricles …   Medical dictionary

  • Pacing — may mean:* Pacing, a horse gait used in Standardbred horse races * Pacing, the even spreading out of athletes strength in longer distance track and field and swimming races, so that they don t burn themselves out * Pacing, an advanced… …   Wikipedia

  • Transvenous pacing — Temporary transvenous cardiac pacing, also called endocardial pacing, is a potentially life saving intervention used primarily to correct profound bradycardia. It can be used to treat symptomatic bradycardias that do not respond to transcutaneous …   Wikipedia

  • transthoracic pacing — a temporary method for cardiac pacing in which a hooked pacing lead is attached to the ventricular endocardium by insertion through the chest wall or epigastric area and then the ventricular wall; it is used in emergency situations, usually… …   Medical dictionary

  • Artificial pacemaker — Cardiac resynchronization therapy and CRT (Cardiac Resynchronization Therapy) redirect here. For the device termed a CRT D, see Implanted cardiac resynchronization device. For other uses, see Pacemaker (disambiguation). A pacemaker, scale in… …   Wikipedia

  • Cardioversion — Intervention ICD 9 CM 99.6 MeSH …   Wikipedia

  • Paramedics in the United States — In the United States, a paramedic is the most advanced prehospital medical professional who responds to and treats medical emergencies and trauma outside of an emergency room. Paramedics typically provide treatment on scene and during… …   Wikipedia

  • Defibrillation — View of defibrillator electrode position and placement. Defibrillation is a common treatment for life threatening cardiac arrhythmias, ventricular fibrillation and pulseless ventricular tachycardia. Defibrillation consists of delivering a… …   Wikipedia

  • Dor procedure — In 1985, Vincent Dor, MD, introduced endoventricular circular patch plasty (EVCPP), or the Dor procedure, as a viable method for restoring a dilated left ventricle to its normal, elliptical geometry. The Dor procedure, which uses a circular… …   Wikipedia

  • TCP — T complex protein; therapeutic carepsychosocial; therapeutic class profile; therapeutic continuous penicillin; thienyl cyclohexylpiperidine; thrombocytopenia; tibial hole coronal positioning; tissue culture treated plastic; tissue or tumor… …   Medical dictionary

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”