- Laser Doppler Vibrometer
A Laser Doppler Vibrometer (LDV) is a scientific instrument that is used to make non-contact vibration measurements of a surface. The
laser beam from the LDV is directed at the surface of interest, and the vibration amplitude and frequency are extracted from theDoppler shift of the laser beam frequency due to the motion of the surface. The output of an LDV is generally a continuous analog voltage that is directly proportional to the target velocity component along the direction of the laser beam.Some advantages of an LDV over similar measurement devices such as an
accelerometer are that the LDV can be directed at targets that are difficult to access, or that may be too small or too hot to attach a physicaltransducer . Also, the LDV makes the vibration measurement without mass-loading the target, which is especially important forMEMS devices.Principles of Operation
A vibrometer is generally a two beam laser
interferometer that measures the frequency (or phase) difference between an internal reference beam and a test beam. The most common type of laser in an LDV is theHelium-Neon laser [ [http://www.polytec.com Polytec, Inc.] ] , althoughlaser diodes [ [http://www.metrolaserldv.com MetroLaser, Inc.] ] ,fiber lasers , and are also used. The test beam is directed to the target, and scattered light from the target is collected and interfered with the reference beam on aphotodetector , typically aphotodiode . Most commercial vibrometers work in aheterodyne regime by adding a known frequency shift (typically 30-40 MHz) to one of the beams. This frequency shift is usually generated by aBragg Cell , or Acousto-Optic Modulator.A schematic of a typical laser vibrometer is shown below. The beam from the laser, which has a frequency fo, is divided into a reference beam and a test beam with a
beamsplitter . The test beam then passes through the Bragg Cell, which adds a frequency shift fb. This frequency shifted beam then is directed to the target. The motion of the target adds a Doppler shift to the beam given by fd = 2*v(t)*cos("alpha"), where v(t) is the velocity of the target as a function of time, and "alpha" is the angle between the laser beam and the velocity vector.Light scatters from the target in all directions, but some portion of the light is collected by the LDV and reflected by the beamsplitter to the photodetector. This light has a frequency equal to fo + fb + fd. This scattered light is combined with the reference beam at the photo-detector. The initial frequency of the laser is very high (> 1014 Hz), which is higher than the response of the detector. The detector does respond, however, to the
beat frequency between the two beams, which is at fb + fd (typically in the tens of MHz range).The output of the photodetector is a standard
frequency modulated (FM) signal, with the Bragg Cell frequency as the carrier frequency, and the Doppler shift as the modulation frequency. This signal can be demodulated to derive the velocity vs. time of the vibrating target.Applications
LDVs are used in a wide variety of scientific, industrial, and medical applications. Some examples are provided below:
* Aerospace - LDVs are being used as tools in non-destructive inspection of aircraft components ["Kilpatrick, James M. and Markov, Vladimir, "Matrix laser vibrometer for transient modal imaging and rapid nondestructive," 8th International Conference on Vibrtaion Measurements by Laser Techniques," SPIE 7098, Ancona, Italy (2008)] .
* Acoustic - LDVs are standard tools for speaker design, and have also been used to diagnose the performance of musical instruments [ [http://findarticles.com/p/articles/mi_qa4075/is_/ai_n19511967, Bissinger, George. and Oliver, David ,"3-D Laser Vibrometry on Legendary Old Italian Violins," Sound and Vibration, July 2007] ] .
* Automotive - LDVs have been used extensively in many automotive applications [http://www.polytec.com/usa/158_2081.asp] , such as structural dynamics, brake diagnostics, and quantification ofNoise,_Vibration,_and_Harshness (NVH).
* Biological - LDVs have been used for diverse applications such as eardrum diagnostics [Huber, Alexander M ,"Evaluation of Eardrum Laser Doppler Interferometry as a Diagnostic Tool," Journal of Comparative Physiology A 111(3):501-507, March 2001] and insect communication [Fonseca, P.J. and Popov, A.V.,"Sound radiation in a cicada: the role of different structures," Volume 175, Number 3, September, 1994, p. 349-361] .
* Calibration - Since LDVs measure motion that can be calibrated directed to the wavelength of the light, there are frequently used to calibrate other types of transducers [Sutton, C. M., Accelerometer Calibration by Dynamic Position Measurement Using Heterodyne Laser Interferometry, Metrologia 27, 133-138, 1990] .
* Hard Disk Drive Diagnostics - LDVs have been used extensively in the analysis of hard disk drives, specifically in the area of head positioning [ [http://books.google.com/books?id=wEnrTMn90B4C&output=html, Mamun, A.A. et al, "Hard Disk Drive: Mechatronics and Control", ISBN 0849372534, 2007] ] .
* Landmine detection - LDVs have shown great promise in the detection of buried landmines. The technique uses an audio source such as a loudspeaker to excite the ground, causing the ground to vibrate a very small amount, and then the LDV to measure the ground vibration. Areas above a buried mine show an enhanced ground velocity at the resonance frequency of the mine-soil system. Mine detection with single-beam scanning LDVs [Xiang, Ning and Sabatier, James M., "Land mine detection measurements using acoustic-to-seismic coupling," SPIE Vol. 4038 Detection and Remediation Technologies for Mines and Minelike Targets V, p.645-655 (2000).] , an array of LDVs [Burgett, Richard D. et al., "Mobile mounted laser Doppler vibrometer array for acoustic landmine detection," SPIE Vol. 5089 Detection and Remediation Technologies for Mines and Minelike Targets VIII (2003).] , and multi-beam LDVs [Lal, Amit K. et al., "Advanced LDV instruments for buried landmine detection," SPIE Vol. 6217 Detection and Remediation Technologies for Mines and Minelike Targets XI (2006).] has been demonstrated.Types of Laser Doppler Vibrometers
* Single Point Vibrometers - This is the most common type of LDV. Vendors include Polytec, MetroLaser, B&K, Brimrose, and Piezojena.
* Scanning Vibrometers - A scanning LDV adds a set of X-Y scanning mirrors, allowing the single laser beam to be moved across the surface of interest.
* 3-D Vibrometers - A standard LDV measures the velocity of the target along the direction of the laser beam. To measure all three components of the target's velocity, a 3-D Vibrometer measures a location with three independent beams, which strike the target from three different directions. This allows a determination of the complete in-plane and out-of-plane velocity of the target.
* Rotational Vibrometers - A rotational LDV is used to measure rotational or angular velocity.
* Differential Vibrometers - A differential LDV measures the out-of-plane velocity difference between two locations on the target.
* Multi-Beam Vibrometers - A multi-beam LDV measures the target velocity at several locations simultaneously.References